Zimmer, Inc.: Bigliani/Flatow The Complete Solution Fukuda Retractors Large Nonsterile 4305-19-40. Used in retracting the humeral shaft posteriorly and exposing the entire glenoid surface.

Recall number
Z-1620-2013
Recalling firm
Zimmer, Inc.
Product
Bigliani/Flatow The Complete Solution Fukuda Retractors Large Nonsterile 4305-19-40. Used in retracting the humeral shaft posteriorly and exposing the entire glenoid surface.
Status
Terminated
Initiated
2013-05-23
Posted
2013-07-01
Terminated
2014-12-18
Reason
This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included. Zimmer, Inc. is initiating a recall of small and large Fukuda-type retractors due to fractures occurring at the grooves around the perimeter of the blades while being used during surgery.
Root cause
Device Design
Action
In 2010, Zimmer initiated a recall of the Bigliani/Flatow Fukuda Retractors due to fractures occurring at the grooves around the perimeter of the blades. Zimmer is expanding this recall to include seven lots that were not previously included. On 5/23/13, an Urgent Medical Device Recall notification letter went to all consignees explaining the recall expansion and risk associated with the recall. All consignees are to locate and quarantine the device immediately. Please return the product along with the completed inventory certification form to Zimmer. If you have any questions regarding this recall, call Zimmer at 877-946-2761.
Quantity
722 units (includes 2010 market withdrawal)
Distribution
Worldwide Distribution-USA (nationwide) including the states of NY, MI, MN, OH, IN, IL, VA, MD, PA, WA, KY, CT, RI, NH, TX, MT, and CA and the countries of Chile, China, South Korea, and Australia.
Lot, serial or model codes
Item Number: 00-4305-019-40 with Lot Numbers 61588143 and 61612276. 2010 Market Withdrawal: Item #:430501940 Lot #'s: 60768814, 60806551, 60893241, 60946820, 60999665, 61031668, 61110808, 61170678, 61205673, 61244194, 61392456, 61430127
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register