Bio-Rad Laboratories: Bio-Rad's Monolisa (TM) Anti-HBs EIA test kit, Labeled as, "MONOLISA TM Anti-HBs EIA 25220; MONOLISA TM Anti-HBs Calibrator Kit 25219... Bio-Rad Laboratories Redmond, WA 98052" For the qualitative detection of antibody to Hepatitis B Surface Antigen (anti-HBs) in Human Serum and EDTA or citrated plasma.

Recall number
Z-1620-2010
Recalling firm
Bio-Rad Laboratories
Product
Bio-Rad's Monolisa (TM) Anti-HBs EIA test kit, Labeled as, "MONOLISA TM Anti-HBs EIA 25220; MONOLISA TM Anti-HBs Calibrator Kit 25219... Bio-Rad Laboratories Redmond, WA 98052" For the qualitative detection of antibody to Hepatitis B Surface Antigen (anti-HBs) in Human Serum and EDTA or citrated plasma.
Status
Terminated
Initiated
2008-08-11
Posted
2010-05-20
Terminated
2010-05-20
Reason
The package insert provided with this kit contains a typographical error, giving incorrect directions to prepare a Working Conjugate Solution. If these directions are followed, it would result in a 1:6 dilution and invalidate the assay.
Root cause
Packaging process control
Action
All domestic customers were notified via an Urgent Product Correction customer letter sent by DHL or FedEx on 8/11/08. Enclosed with the letter was a corrected package insert. The Canadian subsidiary was notified via e-mail on 8/11/08. The letter advised customers of the correct wording, and advised that a new corrected package insert was enclosed and to make note of the correction in the laboratory procedures. Questions should be directed to Bio-Rad Laboratories Technical Support at 1-800-224-6723.
Quantity
795 kits
Distribution
Worldwide Distribution -- USA and Canada.
Lot, serial or model codes
Lot #'s: 288BEE and 056CEE
510(k) numbers
not on file
PMA numbers
["P050048"]
Product code
LOM
Firm FEI
3008272466
Firm city
Redmond
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register