Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Tibial Alignment Handle Triathlon Instruments CAT #: 6541-2-807 Nav Tibial Alignment Handle Triathlon Instruments CAT #; 6541-2-808 Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 The tibial alignment handles Cat# 6451-2-807 and 6541-2-808 are part of the Triathlon Knee Instruments set. The alignment handle is used after the tibial resection portion of a total knee arthroplasty procedure. The alignment handle helps holding and orienting the tibial template and trial, which are used to estimate the size and assess the orientation of the final implant. The implant will provide the new bearing surface of the knee, replacing the damaged joint.

Recall number
Z-1619-2012
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Tibial Alignment Handle Triathlon Instruments CAT #: 6541-2-807 Nav Tibial Alignment Handle Triathlon Instruments CAT #; 6541-2-808 Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 The tibial alignment handles Cat# 6451-2-807 and 6541-2-808 are part of the Triathlon Knee Instruments set. The alignment handle is used after the tibial resection portion of a total knee arthroplasty procedure. The alignment handle helps holding and orienting the tibial template and trial, which are used to estimate the size and assess the orientation of the final implant. The implant will provide the new bearing surface of the knee, replacing the damaged joint.
Status
Terminated
Initiated
2011-10-10
Posted
2012-05-17
Terminated
2014-09-10
Reason
Stryker received a Product Experience Report (PER) that a swivel block was improperly oriented on a Triathlon Tibial Alignment Handle, catalog no. 6541-2-807.
Root cause
Device Design
Action
Stryker sent an Urgent Product Field Action letter dated October 10, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact their Stryker Sales Representative to arrange for return of the affected product . Customers were asked to fax back the attached Poroduct Field Action Acknowledgement Form as soon as possible to 201-831-6069. For any questions customers were instructed to call 201-972-2100. For questions regarding this recall call 201-831-3523.
Quantity
6,126 Units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Asia Pacific, Canada, Europe, Middle East and Africa (EMEA), and Japan
Lot, serial or model codes
Catalog No.: 6541-2-807 N1A0l, N1K07 N1K14 N1N64 N1W04 N2A11 N2H10 N2L16 N2N04 N2V25 N2V26 N3E08 N3H12 N3H29 N3K02 N3K56 N3L13 N3S33 N3S39 N3T78 N3V21 N3W18 N3W29 N3W47 NYC02 NYC10 NYL09 NZC10 NZH09 NZL08 NZM01 NZS08 NZV09 customer Parts associated with cat. no. 6540-2-807 I-K2072AH00 - Triathlon Tibial Align. Handle with Secondary Lock I-K2336AH00 - Tibial Alignment Handle with Lock I-K2021TA00 - Triathlon Alignment Handle with Navigation Post I-K2059KT00 - Triathlon Tibia Alignment Handle I-K3254TA00 - Triathlon Tibial Alignment Handle with Lock I-K3251AH00 - Triathlon Tibial Alignment Handle for Sizes 1-3 I-K3113TH00 - Triathlon Tibial Alignment Handles (Augment) I-K2980 - Triathlon Tibia Alignment Handle I-K2954AH00 - Triathlon Tibia Alignment Handle I-K2916TA00 - Triathlon Tibial Alignment Handle I-K2797AH00 - Tibial Alignment Handle with no swivel Catalog No. 6541-2-8089 N1K14 N1N63 N1W08 N3K50 N3M43 N3T61 N3V56 N4E36
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register