Skip to the content

FDA Medical Device Recalls (openFDA)

I2s Micro Implantable Systems Llc: Inserter Wand Holder, Non-Sterile, Serial # 4273-0000, Blackrock Microsystems, Salt Lake City, UT / or / Cyberkinetics Neurotechnology Systems, Inc., Salt Lake City, UT. Intended use: The Inserter Wand Holder serves two key functions. It firmly secures the Inserter Wand to prevent recoil during the NeuroPort Array insertion process, and it facilitates precision alignment of the Inserter Wand with respect to the back of the NeuroPort Array just prior to pneumatic impulse insertion.

Recall number
Z-1619-2010
Recalling firm
I2s Micro Implantable Systems Llc
Product
Inserter Wand Holder, Non-Sterile, Serial # 4273-0000, Blackrock Microsystems, Salt Lake City, UT / or / Cyberkinetics Neurotechnology Systems, Inc., Salt Lake City, UT. Intended use: The Inserter Wand Holder serves two key functions. It firmly secures the Inserter Wand to prevent recoil during the NeuroPort Array insertion process, and it facilitates precision alignment of the Inserter Wand with respect to the back of the NeuroPort Array just prior to pneumatic impulse insertion.
Status
Terminated
Initiated
2010-03-26
Posted
2010-05-20
Terminated
2010-05-25
Reason
Component of cranial array may not allow sufficient tightening to maintain inserter position.
Root cause
Device Design
Action
The firm, Blackrock Microsystems, sent a "RECALL" letter dated March 26, 2010, to all customers. The letter describes the product, problem and actions to be taken by the customers. The customers were instructed to discontinue use of all affected devices and promptly return any inventory, perform a physical account of your inventory for the recalled product and email a list of all serial numbers to Andrew Gotshalk at agotshalk@blackrockmicro.com, and to please email Andrew Gotshalk even if you do not have any inserter wand holders or have destroyed or removed from use the inserter wand holders. Blackrock will rework or replace all Inserter Wand Holders that is returned under this recall. If you have any questions regarding this recall please call 801-582-5533 x 275.
Quantity
22 units
Distribution
Nationwide distribution: CA, MA, NY, RI, and UT.
Lot, serial or model codes
Serial # 4273-0001 through 4273-0061
510(k) numbers
K042384
PMA numbers
not on file
Product code
GWL
Firm FEI
3007323246
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source