Karl Storz Endoscopy: Karl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers: 1. 8403xx series, which includes: 8403AX, AXC, BX, BXC, DXC, EXC, GXC, HX, HXK, HXP, KXC, MXC 2. 8401xx series, which includes: 8401AX, AXC, BX, BXC, DXC, GXC, HX, HXP, KXC. Product Usage: Video Laryngoscopes are used to visualize the airways and vocal cords during endotracheal intubation and for the inspection and examination of the upper respiratory tract. Video Laryngoscopes are designed for transient use in invasive procedures through a body orifice.

Recall number
Z-1618-2020
Recalling firm
Karl Storz Endoscopy
Product
Karl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers: 1. 8403xx series, which includes: 8403AX, AXC, BX, BXC, DXC, EXC, GXC, HX, HXK, HXP, KXC, MXC 2. 8401xx series, which includes: 8401AX, AXC, BX, BXC, DXC, GXC, HX, HXP, KXC. Product Usage: Video Laryngoscopes are used to visualize the airways and vocal cords during endotracheal intubation and for the inspection and examination of the upper respiratory tract. Video Laryngoscopes are designed for transient use in invasive procedures through a body orifice.
Status
Terminated
Initiated
2019-12-06
Posted
not on file
Terminated
2022-07-01
Reason
The firm has become aware that there is a potential that the glass cover at the distal end of the camera/led lighting on the video laryngoscope could become loose during reprocessing or use. If this occurs during usage, it may result in a delay in procedure and/or require removal from body.
Root cause
Labeling Change Control
Action
On 12/6/2019 a "Urgent Medical device Correction" letter was sent to affected customers via mail. In addition, to providing information on the recall, the firm asked customers to take the following actions: 1. Please forward this Safety Information to the appropriate departments in your organization. 2. Please train your employees on the new Instructions for Use, especially chapter 5.1. 3. Please complete the attached acknowledgement form and return it to us. 4. If you have any questions about this communication, or needs additional copies of this Instruction for Use, please contact our Technical Support at 800-421-0837, Ext. 5350, Option 5.
Quantity
27,165 scopes
Distribution
Worldwide distribution - US Nationwide and country of Canada.
Lot, serial or model codes
All Lots
510(k) numbers
not on file
PMA numbers
not on file
Product code
EQN
Firm FEI
3007475226
Firm city
El Segundo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register