Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Triathlon Revision Instruments; 1,2,7,8 TGC Upper Tray; Non Sterile, Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430. For total knee arthroplasty procedures involving Triathlon Total Stabilizer components. The trays sterilize and transport instruments needed for such procedures.

Recall number
Z-1618-2009
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Triathlon Revision Instruments; 1,2,7,8 TGC Upper Tray; Non Sterile, Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430. For total knee arthroplasty procedures involving Triathlon Total Stabilizer components. The trays sterilize and transport instruments needed for such procedures.
Status
Terminated
Initiated
2009-05-15
Posted
2009-07-15
Terminated
2010-04-05
Reason
The screen print for the Valgus Adapter catalog numbers on the Triathlon Trial Cutting Guide Instrument Trays is incorrect.
Root cause
Process change control
Action
Stryker Orthopaedics issued an "Urgent Product Correction" notice dated May 15, 2009 via Federal Express addressed to Branch/Agency Manager/Quality Contact informing them of the affected device. The letter outlined instructions for correction and requested consignees to complete and return a Product Correction Acknowledgment Form by fax to 1-201-831-6069. Questions can be directed to Colleen O'Meara, Manager, Divisional Regulatory Reporting at 201-831-5970.
Quantity
27
Distribution
Worldwide Distribution -- US and Canada.
Lot, serial or model codes
Catalog Numbers: 6543-8-014, 6543-8-015.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JWH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register