Reason
IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling
Root cause
Under Investigation by firm
Action
Omincell notified Customers via phone on 3/27/2023 and by Urgent Field Action: Mandatory Correction Of Product Labeling Ivx Station (Limited Release) Letter on 3/31/2023 via email. Letter states reason for recall, health risk and action to take: We will analyze specific devices at your Limited Release sites. Upon completion of our analysis updated labeling will be provided and will include the revised dose accuracy range.
Quantity
6 units
Distribution
AL, NY
Lot, serial or model codes
UDI-DI: 00860006285005 Serial Numbers: IVXS000014, IVXS000015, IVXS000016, IVXS000017, IVXS000018. IVXS000019
510(k) numbers
not on file
PMA numbers
not on file
Product code
NEP
Firm FEI
3011278888
Firm city
Warrendale
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source