C.R. Bard, Inc.: Arctic Sun Temperature Management System, 5000, 100-120V and 220-240V The product is a thermal regulating system indicated for monitoring and controlling patient temperature
- Recall number
- Z-1617-2016
- Recalling firm
- C.R. Bard, Inc.
- Product
- Arctic Sun Temperature Management System, 5000, 100-120V and 220-240V The product is a thermal regulating system indicated for monitoring and controlling patient temperature
- Status
- Terminated
- Initiated
- 2015-11-09
- Posted
- 2016-05-06
- Terminated
- 2017-06-06
- Reason
- The product may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery responsible for maintaining the system close and Static Random Access Memory (SRAM). This premature coin cell battery drainage could render the device unresponsive upon System startup.
- Root cause
- Unknown/Undetermined by firm
- Action
- BARD sent an Urgent Medical Product Recall to all affected consignees on November 9, 2015. The letter indicates that a technician will contract the firm for the evaluation and potential replacement of the affected electronic components. The technician will fill out the Field Action Effectiveness Check Form after evaluation. Customers with questions were instructed to call 1-770-784-6471 or email bmd@crbard.com. For questions regarding this recall call 770-784-6220.
- Quantity
- 2613
- Distribution
- Worldwide Distribution to AL, AK, AZ, AR, CA, CO, CT, DEL, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NH, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and Internationally to Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Finland, France, Germany, Gibraltar, Greece, Guam, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom
- Lot, serial or model codes
- Product Codes - 5000-00-00, 5000-01-01, 5000-01-02, 5000-01-03, 5000-01-04, 5000-01-05
- 510(k) numbers
- ["K101092"]
- PMA numbers
- not on file
- Product code
- DWJ
- Firm FEI
- 1018233
- Firm city
- Covington
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28