FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree and the gastrointestinal tract Models: NA-U401SX-4022, NA-U401SX-4022-A
- Recall number
- Z-1616-2023
- Recalling firm
- Olympus Corporation of the Americas
- Product
- To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree and the gastrointestinal tract Models: NA-U401SX-4022, NA-U401SX-4022-A
- Status
- Open, Classified
- Initiated
- 2023-03-31
- Posted
- 2023-05-19
- Terminated
- not on file
- Reason
- Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.
- Root cause
- Employee error
- Action
- Olympus issued Urgent Medical Medical Device Recall letter on 3/31/23 to Endoscopy Department, Pulmonology Department; Risk Management. . Letter states reason for recall, health risk and action to take: 1. Immediately locate any affected products listed in this communication and quarantine these products. 2. Contact our Olympus Customer Service Team at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. " Go to https://olympusamerica.com/recall and enter the recall number 0422 4. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints, including patient injuries and adverse events, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, to report complaints
- Quantity
- 1297 (Boxes; 5 per Box)
- Distribution
- Natiowide Foreign: Canada, Germany
- Lot, serial or model codes
- UDI: 04953170388286 Lot numbers: KR248740 KR248746 KR252863 KR253666 KR253681 KR260854 KR260865 KR260873 KR260875 KR260876 KR260877 KR260896 KR260917 KR260937
- 510(k) numbers
- ["K160098"]
- PMA numbers
- not on file
- Product code
- FCG
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28