FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: ViziShot 2 Single Use Aspiration Needle, 21G- To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree and the gastrointestinal tract. Model: NA-U401SX-4021, NA-U401SX-4021-A

Recall number
Z-1615-2023
Recalling firm
Olympus Corporation of the Americas
Product
ViziShot 2 Single Use Aspiration Needle, 21G- To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree and the gastrointestinal tract. Model: NA-U401SX-4021, NA-U401SX-4021-A
Status
Open, Classified
Initiated
2023-03-31
Posted
2023-05-19
Terminated
not on file
Reason
Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.
Root cause
Employee error
Action
Olympus issued Urgent Medical Medical Device Recall letter on 3/31/23 to Endoscopy Department, Pulmonology Department; Risk Management. . Letter states reason for recall, health risk and action to take: 1. Immediately locate any affected products listed in this communication and quarantine these products. 2. Contact our Olympus Customer Service Team at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. " Go to https://olympusamerica.com/recall and enter the recall number 0422 4. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints, including patient injuries and adverse events, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, to report complaints
Quantity
2058 (Boxes; 5 per Box)
Distribution
Natiowide Foreign: Canada, Germany
Lot, serial or model codes
UDI: 04953170388248 Lot numbers: KR243729 KR248648 KR248652 KR248673 KR248784 KR248785 KR248808 KR249160 KR251593 KR257313 KR257314 KR264662 KR264697 KR264708 KR264711 KR264726 KR264732 KR264746 KR264751 KR264778 KR264806 KR276770
510(k) numbers
["K160098"]
PMA numbers
not on file
Product code
FCG
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register