FDA Medical Device Recalls (openFDA)
Ortho-Clinical Diagnostics: VITROS Chemistry Products CRP Slides (Reagent), REF/Catalog Number 192 6740 and 809 7990, Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products CRP Slides quantitatively measure C-reactive protein (CRP) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
- Recall number
- Z-1615-2013
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- VITROS Chemistry Products CRP Slides (Reagent), REF/Catalog Number 192 6740 and 809 7990, Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products CRP Slides quantitatively measure C-reactive protein (CRP) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
- Status
- Terminated
- Initiated
- 2013-05-09
- Posted
- 2013-06-27
- Terminated
- 2015-06-12
- Reason
- While performing interval testing Ortho Clinical Diagnostics (OCD) determined that an excess amount of surfactant was unintentionally added to VITROS Chemistry Products CRP Slides during the manufacturing process.
- Root cause
- Employee error
- Action
- Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated May 9, 2013 via FedEx or US Priority Mail to all US customers. Foreign Affiliates were notified by e-mail on May 9, 2013 informing them of the issue. Distributors were instructed to notify their customers of the recall. Customers were instructed to discontinue use, discard the affected product, comp[lete and return the attached Confirmation of Recipt Form, and post this notification by each VITROS System. For questions call Customer Service Center at 1-800-828-6316.
- Quantity
- Domestic: 4,257; Foreign: 923
- Distribution
- Worldwide Disribution - USA Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain and Venezuela.
- Lot, serial or model codes
- GEN 3773, Coating 0766, Lot XXXX (exp. 11/01/2013); GEN 3784, Coating 0767, Lot XXXX (exp. 11/01/2013); and GEN 3785, Coating 0768, Lot XXXX (exp. 10/01/2013 and 11/01/2013). The term GEN and Coating refer to specific segments of the reagent lot number.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- DCK
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28