TomoTherapy Incorporated: TomoTherapy Hi-Art System with software versions 3.0 and 3.1 - Radiation Therapy System. The device is intended to be used as an integrated system for the planning and delivery of intensity modulated radiation therapy.

Recall number
Z-1615-2008
Recalling firm
TomoTherapy Incorporated
Product
TomoTherapy Hi-Art System with software versions 3.0 and 3.1 - Radiation Therapy System. The device is intended to be used as an integrated system for the planning and delivery of intensity modulated radiation therapy.
Status
Terminated
Initiated
2008-02-22
Posted
2008-08-24
Terminated
2008-12-02
Reason
Registration adjustment field numbers may appear on printed pages of the Register Tab as nonsense characters or values different than the numbers shown on the Operator Station (OS) display. Setup will move the couch to the correct offset position. If the couch is manually adjusted to the incorrect value, the patient may not be the correct position during treatment.
Root cause
Software design
Action
Consignees were sent a letter "TomoTherapy Incorporated Urgent Medical Device Correction Field Safety Notice" dated February 22,2008. The letter described the Issue, Product Affected, Cause, Recommended Customer Action and Resolution. If you have not received this letter or have questions, contact the firm as outlined below: North America: 1 866 368 4807 Belgium: 0800 38783 France: 0805 631 565 Germany: 0800 000 7541 Italy: 800 986 399 Netherlands: 0800 0201364 Spain: 800 300049 Switzerland: 0800 001927 United Kingdom: 0808 238 6035 China/CNC (Northern): 10 800 712 1701 China (Southern): 10 800 120 1701 Hong Kong: 800 967912 Japan: 0044 22 132374 Singapore: 800 1204 683 South Korea: 0079 81 4800 7204 All other locations: +1 608 824 2900 or +32 2 400 44 44
Quantity
32
Distribution
Worldwide Distribution --- USA including states of CA ,CT, DC, FL, IL, LA, MD, MI, MO, NE, NJ, NY, SC, TN, VA, WI , and countries of Canada, Germany, Taiwan, and Turkey.
Lot, serial or model codes
10002 10003 10007 10016 10019 10021 10028 10037 10044 10048 10050 10051 10056 10094 10165 10166 10168 10169 10174 10185 10186 10187 10188 10189 10190 10191 10192 10193 10194 10195 10196 10198
510(k) numbers
["K060912"]
PMA numbers
not on file
Product code
IYE
Firm FEI
3003873069
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register