FDA Medical Device Recalls (openFDA)
ROUTE 92 MEDICAL INC: Route 92 Medical 070 Reperfusion System, REF 1003, Rxonly, SterileEO. Used for introduction of interventional devices into the neurovasculature and aspiration of thrombus in ischemic stroke patients.
- Recall number
- Z-1614-2024
- Recalling firm
- ROUTE 92 MEDICAL INC
- Product
- Route 92 Medical 070 Reperfusion System, REF 1003, Rxonly, SterileEO. Used for introduction of interventional devices into the neurovasculature and aspiration of thrombus in ischemic stroke patients.
- Status
- Open, Classified
- Initiated
- 2024-03-08
- Posted
- 2024-04-24
- Terminated
- not on file
- Reason
- Their is the potential for distal tip separation at the proximal marker band of the delivery catheters
- Root cause
- Process control
- Action
- On March 12, 2024 Route 92, issued a Urgent Medical Device recall notification to affected consignees via Mail and EMail. Route 92 asked consignees to take the following actions: Ensure all of your SUMMIT MAX inventory including the affected 070 Reperfusion Systems is in a secure, quarantined location. 2. Complete the Acknowledgement Form in Attachment 2 and return to Route 92 Medical by one of the following methods: Email-FieldActions@route92medical.com/Standard Mail - Attention Line: Tenzing 7 fiedl action, 155 Bovet Road Suite 100, San Mateo, CA 94402 Please return the Acknowledgement Form even if you no longer have the product subject to this recall. 4. Immediately return any affected products with the specific lot numbers listed in Attachment 1 to the address above using the original shipping container or request a new shipping container and pre-printed shipping label by email. 5. Keep this notice until all product subject to this recall has been returned. While processing your returns, please maintain a copy of this notice with the quarantined product subject to this recall and keep a copy for your records. 6. Separately, as part of SUMMIT MAX closure, you will be sending back your remaining SUMMIT MAX inventory if you haven t done so already.
- Quantity
- 19 units
- Distribution
- US: AL AZ CA DE FL GA IA IL IN KY LA MA MD MI MN MO NC ND NJ NY OH OR PA RI SD TN TX UT VA WV OUS: New Zealand
- Lot, serial or model codes
- UDI: 00853799007169/ Lot # 23112702 and 23112901
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- DQY
- Firm FEI
- 3016522967
- Firm city
- San Mateo
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28