Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: Codman Cranial Hand Drill- Indicated when a craniotomy is required for placement of an intracranial pressure (ICP) monitoring device and/or cerebrospinal fluid drainage device. Part Number: 82-6607

Recall number
Z-1614-2023
Recalling firm
Integra LifeSciences Corp.
Product
Codman Cranial Hand Drill- Indicated when a craniotomy is required for placement of an intracranial pressure (ICP) monitoring device and/or cerebrospinal fluid drainage device. Part Number: 82-6607
Status
Open, Classified
Initiated
2023-04-11
Posted
2023-05-19
Terminated
not on file
Reason
Released with discoloration (rust) on the Drill Chuck and inner chuck of Codman Cranial Hand Drill may result in inflammation, local toxicity, systemic toxicity, sensitivity, revision surgery, procedural delay, inconvenience to user, and/or infection.
Root cause
Under Investigation by firm
Action
Codman issued Urgent Medical Device Recall Letter on 4/11/23 via Fed'X. Letter states reason for recall, health risk and action to take: 1. If you do have units of the affected product listed in Table 1, remove them immediately from service. 2. If you do have affected product, check the box on the enclosed form I do have affected product. Record the total quantity of the affected product that you have. 3. If you do not have affected product, check the box, I do not have affected product. 4. Complete the attached Acknowledgement Form and return to FCA2@integralife.com i. or FAX to 1-609-750-4220. 5. Keep a copy of the form for your records. 6. When your form is received, and it is noted that you have affected product, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. A credit or alternative replacement can be placed for the quantity noted on the form. Receipt of this form ensures that Integra has achieved a level of effectiveness in communicating this information. We recommend you also maintain a copy of this notification and signed copy of the acknowledgement form for your records. Regulatory agencies such as the FDA perform audits of field actions of this nature to verify that customers have been notified and understand the nature of the field action. Should you have any questions regarding these instructions, please contact your Integra Sales Representative or Customer Service: Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com
Quantity
337 units
Distribution
PA Foreign: Australia, Belgium
Lot, serial or model codes
UDI-DI: 10381780520337 Lot Numbers: 6808685, 6563961, 6568711, 6568709, 6568710
510(k) numbers
["K914479"]
PMA numbers
not on file
Product code
GWM
Firm FEI
3003418325
Firm city
Princeton
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register