Monteris Medical Corp: NeuroBlate System SideFire Select 2.2 mm Directional Laser Probe, Sterile EO. Model #s SFS000-01,SFS002-01, SFS122-01, SFS222-01, SFS322-01, SFS422-01, and SFS522-01. NeuroBlate System SideFire Select Directional Laser Probes are part of the NeuroBlate System. The Monteris Medical NeuroBlate System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It also provides real-time thermographic analysis of selected MRI images. The laser delivery probes direct the laser energy to the proper treatment location.
- Recall number
- Z-1614-2016
- Recalling firm
- Monteris Medical Corp
- Product
- NeuroBlate System SideFire Select 2.2 mm Directional Laser Probe, Sterile EO. Model #s SFS000-01,SFS002-01, SFS122-01, SFS222-01, SFS322-01, SFS422-01, and SFS522-01. NeuroBlate System SideFire Select Directional Laser Probes are part of the NeuroBlate System. The Monteris Medical NeuroBlate System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It also provides real-time thermographic analysis of selected MRI images. The laser delivery probes direct the laser energy to the proper treatment location.
- Status
- Terminated
- Initiated
- 2016-03-25
- Posted
- 2016-05-16
- Terminated
- 2017-03-10
- Reason
- Monteris received a report that the sapphire lens on a 2.2SF probe fractured during a LITT procedure, with resulting introduction of C02 into the patient's brain.
- Root cause
- Device Design
- Action
- Monteris Medical Corporation field personnel visited each consignee starting March 25, 2016 and April 1, 2016. The devices were physically removed from each site. Customers from whom the devices were removed will either receive a credit or receive replacements of other models of laser delivery probes.
- Quantity
- 121
- Distribution
- Nationwide Distribution in the states of KS, MO, NC, OH, and VA.
- Lot, serial or model codes
- All unexpired lots of these models are being removed from the market.
- 510(k) numbers
- ["K143457"]
- PMA numbers
- not on file
- Product code
- GEX
- Firm FEI
- 3009970070
- Firm city
- Plymouth
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28