Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics: VITROS Chemistry Products K+ (Potassium) Slides (Reagent), REF/Catalog Number 815 7596, Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products K+ Slides quantitatively measure potassium (K+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Recall number
Z-1614-2013
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS Chemistry Products K+ (Potassium) Slides (Reagent), REF/Catalog Number 815 7596, Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products K+ Slides quantitatively measure potassium (K+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Status
Terminated
Initiated
2013-05-09
Posted
2013-06-27
Terminated
2015-06-12
Reason
While performing interval testing, Ortho Clinical Diagnostics (OCD) determined that a surfactant was unintentionally added to VITROS Chemistry Products K+ Slides during the manufacturing process.
Root cause
Employee error
Action
Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated May 9, 2013 via FedEx or US Priority Mail to all US customers. Foreign Affiliates were notified by e-mail on May 9, 2013 informing them of the issue. Distributors were instructed to notify their customers of the recall. Customers were instructed to discontinue use, discard the affected product, comp[lete and return the attached Confirmation of Recipt Form, and post this notification by each VITROS System. For questions call Customer Service Center at 1-800-828-6316.
Quantity
Domestic: 236,364; Foreign: 8,339
Distribution
Worldwide Disribution - USA Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain and Venezuela.
Lot, serial or model codes
GEN 4102, Coating 0884, Lot XXXX (exp. 06/01/2014) and GEN 4102, Coating 0885, Lot XXXX (exp. 06/01/2014 and 07/01/2014). The term GEN and Coating refer to specific segments of the reagent lot number.
510(k) numbers
["K083680"]
PMA numbers
not on file
Product code
CEM
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register