Olympus Corporation of the Americas: Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code: Catalogue #N5786100

Recall number
Z-1613-2023
Recalling firm
Olympus Corporation of the Americas
Product
Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code: Catalogue #N5786100
Status
Open, Classified
Initiated
2023-04-18
Posted
2023-05-19
Terminated
not on file
Reason
The Single Use Distal Cover MAJ-2315 may cause mucosal injury and/or become detached from the endoscope during use. Consequences of a detached cover could include the risk of aspiration, inhalation, or obstruction that will require urgent removal, and/or burns from an uncovered distal end.
Root cause
Device Design
Action
An URGENT - Medical Device Recall notification letter dated 4/18/23 was sent to customers. Instructions for Use: Correct application and inspection of the distal cover before the procedure is critical. In addition, inspection of the distal end after the procedure is also important. Olympus representatives are available to provide additional support and training. Please contact your local Olympus facility to further discuss. Action Required: Our records indicate that you have purchased or may use the MAJ-2315 Distal Covers with the original design. Olympus requires you to take the following action: 1. Immediately locate and no longer use the MAJ-2315 with the original design. The original design can be identified by the item code (below) or by lot numbers less than H2530. The image below depicts the area where the lot number is identified. The lot number is on the Carton box and Pack. Model: MAJ-2315 Item code: N5786100 Lot number: Lower than H2530 2. Contact our Olympus Customer Service Team at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a) Go to https://olympusamerica.com/recall b) Enter the recall number: 0423 c) Complete the form as instructed. 4. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints, including mucosal injuries, distal cover detachment and adverse events, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, to report complaints. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration
Quantity
309278 pieces
Distribution
US Nationwide.
Lot, serial or model codes
UDI-DI: 04953170403019; Item code: Catalog N5786100 Lots Lower than H2530
510(k) numbers
["K193182"]
PMA numbers
not on file
Product code
FDT
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register