FDA Medical Device Recalls (openFDA)

ROUTE 92 MEDICAL INC: Route 92 Medical, 070 Access System, REF 1004, Rxonly , Sterile EO. Used for introduction of interventional devices into the neurovasculature.

Recall number
Z-1612-2024
Recalling firm
ROUTE 92 MEDICAL INC
Product
Route 92 Medical, 070 Access System, REF 1004, Rxonly , Sterile EO. Used for introduction of interventional devices into the neurovasculature.
Status
Open, Classified
Initiated
2024-03-08
Posted
2024-04-24
Terminated
not on file
Reason
Their is the potential for distal tip separation at the proximal marker band of the delivery catheters
Root cause
Process control
Action
On March 12, 2024 Route 92, issued a Urgent Medical Device recall notification to affected consignees via Mail and EMail. Route 92 asked consignees to take the following actions: Ensure all of your SUMMIT MAX inventory including the affected 070 Reperfusion Systems is in a secure, quarantined location. 2. Complete the Acknowledgement Form in Attachment 2 and return to Route 92 Medical by one of the following methods: Email-FieldActions@route92medical.com/Standard Mail - Attention Line: Tenzing 7 fiedl action, 155 Bovet Road Suite 100, San Mateo, CA 94402 Please return the Acknowledgement Form even if you no longer have the product subject to this recall. 4. Immediately return any affected products with the specific lot numbers listed in Attachment 1 to the address above using the original shipping container or request a new shipping container and pre-printed shipping label by email. 5. Keep this notice until all product subject to this recall has been returned. While processing your returns, please maintain a copy of this notice with the quarantined product subject to this recall and keep a copy for your records. 6. Separately, as part of SUMMIT MAX closure, you will be sending back your remaining SUMMIT MAX inventory if you haven t done so already.
Quantity
117 units
Distribution
US: AL AZ CA DE FL GA IA IL IN KY LA MA MD MI MN MO NC ND NJ NY OH OR PA RI SD TN TX UT VA WV OUS: New Zealand
Lot, serial or model codes
UDI: 00853799007442/ Lot # 23061601 24010306, 23081802, 24011801, 23100601, 24020801, 23102601
510(k) numbers
["K203043"]
PMA numbers
not on file
Product code
QJP
Firm FEI
3016522967
Firm city
San Mateo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register