Blackstone Medical, Inc.: 5.5MM Bone Tap, Modular; Manufactured by: Blackstone Medical, Inc. 1720 Bray Central Drive, McKinney, TX 75069. Intended to created a screw thread in a hole drilled into bone to facilitate placement of a bone screw.
- Recall number
- Z-1612-2010
- Recalling firm
- Blackstone Medical, Inc.
- Product
- 5.5MM Bone Tap, Modular; Manufactured by: Blackstone Medical, Inc. 1720 Bray Central Drive, McKinney, TX 75069. Intended to created a screw thread in a hole drilled into bone to facilitate placement of a bone screw.
- Status
- Terminated
- Initiated
- 2010-04-21
- Posted
- 2010-05-19
- Terminated
- 2010-08-10
- Reason
- Certain lots of the non-cannulated Modular Bone Taps used with the Firebird Spinal Fixation Systems may contain a manufacturing non conformance resulting in the taps creating a single lead screw thread rather than the required dual lead screw thread.
- Root cause
- Nonconforming Material/Component
- Action
- Orthofix issued an "Urgent Product Recall" notification dated April 21, 2010 by USPS certified mail to hospitals and distributors. Consignees were asked to return all affected product to the firm. For additional information, contact the Orthofix Regulatory Affairs Department at 1-973-406-2847.
- Quantity
- 94
- Distribution
- Nationwide and to one location each in Australia and Spain.
- Lot, serial or model codes
- Part Number 52-1025; Lot numbers16380-xxxx and 17162-xxxx
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HWX
- Firm FEI
- 3001238051
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28