Skip to the content

FDA Medical Device Recalls (openFDA)

Blackstone Medical, Inc.: 5.5MM Bone Tap, Modular; Manufactured by: Blackstone Medical, Inc. 1720 Bray Central Drive, McKinney, TX 75069. Intended to created a screw thread in a hole drilled into bone to facilitate placement of a bone screw.

Recall number
Z-1612-2010
Recalling firm
Blackstone Medical, Inc.
Product
5.5MM Bone Tap, Modular; Manufactured by: Blackstone Medical, Inc. 1720 Bray Central Drive, McKinney, TX 75069. Intended to created a screw thread in a hole drilled into bone to facilitate placement of a bone screw.
Status
Terminated
Initiated
2010-04-21
Posted
2010-05-19
Terminated
2010-08-10
Reason
Certain lots of the non-cannulated Modular Bone Taps used with the Firebird Spinal Fixation Systems may contain a manufacturing non conformance resulting in the taps creating a single lead screw thread rather than the required dual lead screw thread.
Root cause
Nonconforming Material/Component
Action
Orthofix issued an "Urgent Product Recall" notification dated April 21, 2010 by USPS certified mail to hospitals and distributors. Consignees were asked to return all affected product to the firm. For additional information, contact the Orthofix Regulatory Affairs Department at 1-973-406-2847.
Quantity
94
Distribution
Nationwide and to one location each in Australia and Spain.
Lot, serial or model codes
Part Number 52-1025; Lot numbers16380-xxxx and 17162-xxxx
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWX
Firm FEI
3001238051
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register