Skip to the content

FDA Medical Device Recalls (openFDA)

Varian Medical Systems Oncology Systems: Eclipse Treatment Planning System, Client 8.1.18 and DCF 8.1.17, model number H48

Recall number
Z-1612-2008
Recalling firm
Varian Medical Systems Oncology Systems
Product
Eclipse Treatment Planning System, Client 8.1.18 and DCF 8.1.17, model number H48
Status
Terminated
Initiated
2008-02-13
Posted
2008-09-23
Terminated
2011-05-17
Reason
A software error causes the wedge accessory calculation to be ignored in the radiotherapy treatment plan.
Root cause
Software design
Action
not on file
Quantity
90 units
Distribution
Product was distributed throughout the USA (in LA, IN, GA, MD, IL, AR, CA, MA, FL, AZ, SC, MO, CO, AL, OK, WA, MI, IN, NY, TX, MS, NE, TN, VA, WI, OR, DC, and NC) and internationally (Japan, Columbia, Canada, India, Czech Republic, Turkey, Germany, Portugal, Venezuela, Australia, South Korea).
Lot, serial or model codes
Serial numbers all starting with H48 (model number): 6509, 6974, 6616, 6688, 6806, 6568, 6530, 6839, 6740, 7178, 6700, 6694, 6505, 6924, 6979, 6715, 6556, 6757, 6820, 6545, 6564, 7027, 6783, 6553, 6760, 6904, 6907, 7037, 6639, 7088, 6702, 6659, 6940, 6802, 6856, 6944, 6841, 6870, 6874, 6928, 6913, 6614, 7009, 6951, 6493, 6758, 7030, 6453, 6704, 6555, 6691, 6914, 6739, 6846, 6685, 6473, 7015, 6875, 6566, 6599, 6818, 7020, 6799, 6804, 6699, 7075, 6901, 6827, 6833, 6681, 6724, 6499, 6698, 6722, 7018, 6782, 6893, 6522, 6937, 6708, 6560, 6977, 6978, 6803, 6961, 6503, 6814, 7029, 6707, 6686
510(k) numbers
["K071873"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register