Skip to the content

FDA Medical Device Recalls (openFDA)

Fujifilm Medical Systems U.S.A., Inc.: FujiMedical Synapse Cardiovascular I ProSolv CardioVascular. A picture Archiving and Communication System Software versions: Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1

Recall number
Z-1611-2015
Recalling firm
Fujifilm Medical Systems U.S.A., Inc.
Product
FujiMedical Synapse Cardiovascular I ProSolv CardioVascular. A picture Archiving and Communication System Software versions: Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1
Status
Terminated
Initiated
2015-02-04
Posted
2015-05-14
Terminated
2018-09-20
Reason
Measurement values in the EMR may not accurately reflect the actual measurement value on the Synapse Cardiovascular clinical report.
Root cause
Software in the Use Environment
Action
FujiMedical Systems North America issued intial recall letter dated February 4, 2014, advising users of the HL7 report exported to the EMR wherein measurement values presented in the Hl7 reports in your EMR may not accurately reflect the actual measurement value on the Synapse Cardiovascular clinical report. Follow-up letters were issued February 12, 2014, and July 24, 2014. A mandatory upgrade to software version 5.1.1/ base, in-place upgrade required. A customer response form attached to acknowledge receipt and arrange for the on site upgrade. For questions regarding this recall call 203-324-2000.
Quantity
30 units
Distribution
Worldwide Distribution - US including CA, FL, IL, IN, MD, MI, MN, MO, NC, NY, OH, OR, PA, PR, TX, VA, WI Foreign: Canada
Lot, serial or model codes
Software versions: Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1
510(k) numbers
["K023112"]
PMA numbers
not on file
Product code
DQK
Firm FEI
1000513161
Firm city
Stamford
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register