Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Brand Name: Multiday lnfusor, Seven Day lnfusor and the Basal/Bolus lnfusor: Single day Infusor 2 ml/h, Single Day INFUSOR 2 ml/h System, Half Day INFUSOR SV 5 ml/h System, Two Day INFUSOR 2 ml/h System, Two day Infusor 2 ml/h, Multiday Infusor 0.5 ml/h, Multiday INFUSOR 0.5ml/h System, Seven Day Infusor 0.5 ml/h, Seven Day INFUSOR 0.5ml/h System, Basal-Bolus Infusor-15min., Basal/Bolus INFUSOR 0.5 x 2 ml/h System with 15 Minute Lockout, Basal-Bolus Infusor Device Listing: D018045 The Multiday lnfusor. Seven Day lnfusor and the BasaVBolus lnfusor are indicated for patients requiring slow, continuous intravenous, intra-arterial, subcutaneous or epidural administration of medications.

Recall number
Z-1610-2013
Recalling firm
Baxter Healthcare Corp.
Product
Brand Name: Multiday lnfusor, Seven Day lnfusor and the Basal/Bolus lnfusor: Single day Infusor 2 ml/h, Single Day INFUSOR 2 ml/h System, Half Day INFUSOR SV 5 ml/h System, Two Day INFUSOR 2 ml/h System, Two day Infusor 2 ml/h, Multiday Infusor 0.5 ml/h, Multiday INFUSOR 0.5ml/h System, Seven Day Infusor 0.5 ml/h, Seven Day INFUSOR 0.5ml/h System, Basal-Bolus Infusor-15min., Basal/Bolus INFUSOR 0.5 x 2 ml/h System with 15 Minute Lockout, Basal-Bolus Infusor Device Listing: D018045 The Multiday lnfusor. Seven Day lnfusor and the BasaVBolus lnfusor are indicated for patients requiring slow, continuous intravenous, intra-arterial, subcutaneous or epidural administration of medications.
Status
Terminated
Initiated
2013-06-07
Posted
2013-06-27
Terminated
2014-06-04
Reason
Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.
Root cause
Device Design
Action
Baxter sent an Urgent Produd Recall notification letter dated June 13, 2013, to all affected consignees via USPS. The letter was accompanied by a Customer Reply Form and Affected Product List (table). The letter stated affected products, problem description, risk to health (or health risks), actions to be taken by customers/users and contact information. The recall notification instructs customers to segregate affected product from current inventory and contact Baxter for instructions to return affected product. Customers with questions were instructed to call Baxter at 1-800-437-5176, For questions regarding this recall call 847-270-5556.
Quantity
601,230 total
Distribution
Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.
Lot, serial or model codes
Product Codes: 2C1071KJP, 2C1073KJP, 2C1075KJP, 2C1080KJP, 2C1082KJP, 2C1955KJP, 2C1976KJ.  Expiration Dates: 06/30/13, 12/31/13, 06/30/14, 12/31/14
510(k) numbers
["K884505"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register