FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corp.: Brand Name: Coiled Tube Infusors (Portable Elastomeric Infusion Systems). Baxter's Infusor SV Elastomeric Infusion Device Device Listing: D020871 Baxter's lnfusors SV Elastomeric Infusion Device is a single-use, disposable elastomeric infusion pump indicated tor patients requiring slow, continuous intravenous, intra-arterial, subcutaneous, or epidural administration of medications at a constant flow rate.
- Recall number
- Z-1609-2013
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Brand Name: Coiled Tube Infusors (Portable Elastomeric Infusion Systems). Baxter's Infusor SV Elastomeric Infusion Device Device Listing: D020871 Baxter's lnfusors SV Elastomeric Infusion Device is a single-use, disposable elastomeric infusion pump indicated tor patients requiring slow, continuous intravenous, intra-arterial, subcutaneous, or epidural administration of medications at a constant flow rate.
- Status
- Terminated
- Initiated
- 2013-06-07
- Posted
- 2013-06-27
- Terminated
- 2014-06-04
- Reason
- Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.
- Root cause
- Device Design
- Action
- Baxter sent an Urgent Produd Recall notification letter dated June 13, 2013, to all affected consignees via USPS. The letter was accompanied by a Customer Reply Form and Affected Product List (table). The letter stated affected products, problem description, risk to health (or health risks), actions to be taken by customers/users and contact information. The recall notification instructs customers to segregate affected product from current inventory and contact Baxter for instructions to return affected product. Customers with questions were instructed to call Baxter at 1-800-437-5176, For questions regarding this recall call 847-270-5556.
- Quantity
- 601,230 total
- Distribution
- Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.
- Lot, serial or model codes
- Product Code: 2C1073KJP Lot Numbers: 08M002, 08N007, 08N029, 08N030, 08N034, 08N069, 08N070, 09A001
- 510(k) numbers
- ["K062457"]
- PMA numbers
- not on file
- Product code
- MEB
- Firm FEI
- 1416980
- Firm city
- Round Lake
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28