Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Vanguard Knee System AS Tibial Bearing / DCM ArCom, 12 MM X 75 MM; Model Number 189082. The product is placed into a foil pouch with protective packaging in the form of a polyurethane bag, and sealed closed (inner barrier). The inner barrier is then placed into a PETG tray and sealed with Tyvek lid stock (outer barrier). The sealed double barrier system is placed into a paperboard carton with an IFU and label/indicator dots applied to the carton.

Recall number
Z-1607-2024
Recalling firm
Biomet, Inc.
Product
Vanguard Knee System AS Tibial Bearing / DCM ArCom, 12 MM X 75 MM; Model Number 189082. The product is placed into a foil pouch with protective packaging in the form of a polyurethane bag, and sealed closed (inner barrier). The inner barrier is then placed into a PETG tray and sealed with Tyvek lid stock (outer barrier). The sealed double barrier system is placed into a paperboard carton with an IFU and label/indicator dots applied to the carton.
Status
Open, Classified
Initiated
2024-03-13
Posted
2024-04-22
Terminated
not on file
Reason
Possible damage to the Vanguard Knee System AS Tibial Bearing caused during the manufacturing process.
Root cause
Process change control
Action
On March 27, 2024 URGENT MEDICAL DEVICE RECALL letters were via Email to Distributors and mail to customers. Customers were instructed to: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. This medical device recall was reported to the U.S. Food and Drug Administration and will be reported to other Competent Authorities, Notified Bodies, and Regulatory Authorities as required. Med Watch Reporting: Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s Med Watch Adverse Event Reporting program either online, by mail, or by fax. Online: www.fda.gov/medwatch/report.htm Mail: Use postage paid, pre-addressed form FDA 3500, available at: www.fda.gov/MedWatch/getforms.htm Fax: 1-800-FDA-0178. Under 21 CFR 803, manufacturers are also required to report any
Quantity
19 units (1 US, 18 OUS)
Distribution
Worldwide distribution - US Nationwide in the state of IN and the countries of Canada, India, Netherlands, and P.R. China.
Lot, serial or model codes
Item Number: 189082; Lot Number: 65966530; UDI-DI: (01)00880304436954(17)280413(10)65966530;
510(k) numbers
["K050222"]
PMA numbers
not on file
Product code
MBH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register