Toshiba American Medical Systems Inc: Toshiba American Medical Systems Celesteion, Item Code: PCA-9000A/2, 9000A/3B, 9000A/2B This system is designed to acquire CT images, PET images, and fusion images (superimposed PET and CT images). The images that can be acquired include planar images (for any position and from any direction) and volume images (3D images) of the whole human body (including the head). The CT images provide morphological information showing the positional relationships between the lesion and adjacent tissues, while the PET images provide functional information based on the uptake of the injected radioisotope.

Recall number
Z-1607-2016
Recalling firm
Toshiba American Medical Systems Inc
Product
Toshiba American Medical Systems Celesteion, Item Code: PCA-9000A/2, 9000A/3B, 9000A/2B This system is designed to acquire CT images, PET images, and fusion images (superimposed PET and CT images). The images that can be acquired include planar images (for any position and from any direction) and volume images (3D images) of the whole human body (including the head). The CT images provide morphological information showing the positional relationships between the lesion and adjacent tissues, while the PET images provide functional information based on the uptake of the injected radioisotope.
Status
Terminated
Initiated
2016-04-24
Posted
2016-05-04
Terminated
2017-02-22
Reason
Potential degradation of detector performance.
Root cause
Device Design
Action
Toshiba sent an Urgent Medical Device Correction letter dated April 24, 2016 to all affected customers. The letter informed customers that Toshiba America Medical Systems is bringing attention to a potential problem with the Celestion PET/CT System. The letter informed the customers of a potential degradation of detector performance and informed the customers of the problems and the corrective actions to be taken. Customers were instructed to complete and return the attached form and fax it to (877) 349-3054 or be sent via email to regulatoryaffairs@toshiba.com. Customers with questions were instructed to call (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968.
Quantity
3 units
Distribution
US Distribution to WI and NV.
Lot, serial or model codes
Serial No. 3BA1572002, 3BA1562001, 2881492002
510(k) numbers
["K140651"]
PMA numbers
not on file
Product code
KPS
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register