Fujirebio Diagnostics, Inc.: MESOMARK Humanitarian Device. Authorized by Federal Law for use as an aid in monitoring of patients diagnosed with mesothelioma. Each kit contains 96 tests. MESOMARK is an enzyme linked immunosorbent assay (ELISA) for the quantitative measurement of Soluble mesothelin Related Peptides (SMRP) in human serum. Measurement of SMRP may aid in the monitoring of patients diagnosed with epithelioid or biphasic mesothelioma. MESOMARK values must be interpreted in conjunction with all other available clinical and laboratory data.

Recall number
Z-1606-2011
Recalling firm
Fujirebio Diagnostics, Inc.
Product
MESOMARK Humanitarian Device. Authorized by Federal Law for use as an aid in monitoring of patients diagnosed with mesothelioma. Each kit contains 96 tests. MESOMARK is an enzyme linked immunosorbent assay (ELISA) for the quantitative measurement of Soluble mesothelin Related Peptides (SMRP) in human serum. Measurement of SMRP may aid in the monitoring of patients diagnosed with epithelioid or biphasic mesothelioma. MESOMARK values must be interpreted in conjunction with all other available clinical and laboratory data.
Status
Terminated
Initiated
2010-07-29
Posted
2011-03-08
Terminated
2012-07-05
Reason
presence of material attached to the bottom of some wells in places visible to the naked eye, which caused an overestimation of the sample.
Root cause
Other
Action
Notification dated 7/7/10 was sent to the customer via e-mail and FedEx Priority Overnight. FUJIREBIO Diagnostics, Inc. sent an URGENT PRODUCT CORRECTION NOTIIFICATION letter dated July 7, 2010, to all affected customers via e-mail and FedEx Priority Overnight. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue using MESOMARK kit. Lot 57K04510 and discard any remaining inventory. Complete and return the attached Confirmation of Receipt form upon receipt of this notification, no later than July 30, 2010. Notify your ordering physician or health care provider as appropriate. For questions regarding this recall call 877-861-7246 or 610-240-3800.
Quantity
42 kits
Distribution
Nationwide Distribution including UT
Lot, serial or model codes
Lot: 57K04510, Item # 801-905, Expiration Date: December 14, 2010 and Lot: 57K04409, Item# 801-905, Expiration Date: June 15, 2010
510(k) numbers
not on file
PMA numbers
not on file
Product code
OAW
Firm FEI
2521625
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register