Respironics California Inc: Respironics BiPAP Focus Noninvasive Ventilator System. The BiPAP Focus is a non-invasive, pressure support ventilator used to augment the breathing of patients suffering from acute or chronic respiratory insufficiency, respiratory failure or to maintain airway patency and provide respiratory support to patients who experience obstructive sleep apnea. Nasal and full face masks are used with this device. The BiPAP Focus is intended for use in acute, sub-acute and intra-hospital transport settings.

Recall number
Z-1606-2009
Recalling firm
Respironics California Inc
Product
Respironics BiPAP Focus Noninvasive Ventilator System. The BiPAP Focus is a non-invasive, pressure support ventilator used to augment the breathing of patients suffering from acute or chronic respiratory insufficiency, respiratory failure or to maintain airway patency and provide respiratory support to patients who experience obstructive sleep apnea. Nasal and full face masks are used with this device. The BiPAP Focus is intended for use in acute, sub-acute and intra-hospital transport settings.
Status
Terminated
Initiated
2008-09-22
Posted
2009-07-07
Terminated
2010-07-06
Reason
Power Supply failures have occurred on some ventilators. A discrepancy between the power supply strain relief clip and the power cord connector may allow unwanted movement of the cord and the unit's power supply. A disconnection or intermittent electrical connection between the power cord and the unit's power supply during operation or charging can result in transient power surges, which may cau
Root cause
Component design/selection
Action
Respironics California, Inc. issued an "Attention - Medical Device Field Correction" letter addressed to the Customer/Distributor dated September 22, 2008. The letter informed consignees of the affected device and instructions for removal, device replacement and return of the Business Response form by fax to 1-760-918-0169. Further question may be addressed to Respironics US Customer Service at 1-9800-345-6443 extension 5 or Respironics International Customer Service at 1-724-387-4000.
Quantity
2,270 Power Cords
Distribution
Worldwide to Argentina, Australia, Belgium, Brazil, Cambodia, Canada, Colombia, Denmark, Ecuador, Egypt, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Libya, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Portugal, Puerto Rico, Russian Fed., Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, United States and Venezuela.
Lot, serial or model codes
Model Numbers: 8-400169-00, 7-400076-00, 8-400179-00, 8-400258-00, 8-400259-00, 8-400260-00, 8-400261-00 and 8-400216-00.
510(k) numbers
["K053168"]
PMA numbers
not on file
Product code
MNS
Firm FEI
2518422
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register