Philips Medical Systems North America Co. Phillips: Philips BV Pulsera, Rel. 2.3, 718095. Mobile C-Arm X-Ray systems offering Radiographic and Fluoroscopic techniques in a wide variety of applications.

Recall number
Z-1605-2011
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Philips BV Pulsera, Rel. 2.3, 718095. Mobile C-Arm X-Ray systems offering Radiographic and Fluoroscopic techniques in a wide variety of applications.
Status
Terminated
Initiated
2010-12-14
Posted
2011-03-08
Terminated
2012-02-28
Reason
The fixing strap of the belt holding the CBX Laser to the Image Intensifier may be detached during a procedure when used with the BV Libra and the BV Pulsera/Endura Systems.
Root cause
Device Design
Action
On 01/06/11, Philips sent an URGENT Device Correction Notice, dated 01/05/2011, to their consignees. The consignees were informed that the fixing strap of the Laser Aiming Device Image Intensifier may inadvertently detach when used with BV Libra, BV Endura Rel. 2, and BV Pulsera Rel 2.3 Mobile X-Ray systems. Specifically the draw latch of the fixing strap holding the CBX Laser to the Image Intensified can be accidentally released during a procedure when used with the above Mobile X-Ray systems. While waiting for Philips to implement their corrective action, the consignees are advised to fix the closed latch body of the draw latch to the metal band using a Tie Wrap. Without securing the latch with a Tie Wrap, the device should not be used. The consignees can contact Philips Call Center at 1-800-722-9377, select 5, and reference FCO 71800035 with any questions about this recall.
Quantity
214 units
Distribution
Nationwide Distribution -- AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Washington, DC.
Lot, serial or model codes
Site Numbers:  541558, 543462, 543465, 546516, 546531, 546786, 548972, 549275, 549674, 551422, 551444, 551536, 552246, 552452, 552642, 553590, 554473, 554778, 555172, 555727, 555734, 556193, 556206, 556207, 556887, 556920, 557242, 557528, 557570, 557711, 558004, 558232, 558418, 558531, 558632, 558698, 558867, 558925, 558966, 558967, 559322, 559533, 559655, 559658, 559679, 718093, 718101, 41443721, 41444449, 41444577, 41445484, 41445534, 41445599, 41445718, 41445794, 41445895, 41446243, 41446285, 41566245, 41577033, 41587498, 41611719, 41620935, 41620966, 41621991, 41630562, 41645797, 41657683, 41725782, 41748088, 41748089, 41757996, 41780384, 41781354, 41793882, 41885442, 41898013, 41936644, 41938193, 41957602, 41980506, 41986348, 41986353, 41986793, 41987067, 42028669, 42057341, 42182116, 42451705, 42454796, 42474222, 42474223, 42474572, 42474590, 42474663, 42474665, 42474666, 42485796, 42573106, 42573206, 42573399, 42602299, 42616997, 42624941, 42638526, 42657622, 42672459, 42673598, 42677649, 42678321, 42678333, 42678370, 42792818, 42792938, 42859118, 42861795, 42865605, 42869019, 42878225, 42878264, 42878386, 42878440, 42879150, 42879151, 42879230, 42879779, 42891959, 42943097, 42967767, 42968938, 43121942, 43188697, 43246451, 43443832, 43444160, 43444432, 43627105, 43670825, 43708894, 43721186, 43721760, 43721768, 43901634, 43901635, 43972187, 44038499, 44066851, 44235029, 44267704, 44280799, 44330157, 44331403, 44468017, 44489300, 44618133, 45003856, 45030262, 45142308, 45147848, 45147983, 45149689, 45167263, 45168542, 45170408, 45298834, 45303038, 45354311, 45419191, 45428659, 45428664, 45567407, 45586672, 45613075, 45632498, 45641330, 45641331, 47171051, 47349848, 47403000, 47405559, 47420473, 47420474, 47480674, 47539379, 49025829, 49316313, 49368038, 49521186, 49580142, 49753197, 49755484, 49755485, 49875328, 50093136, 50103844, 50110372, 50230731, 50264654, 50305994, 50467164, 50498513, 50829949, 50829955, 51232777, 51240352, 51240364, 51240371, 51240383, 51240386, 51674316, 51702012, 51834968, 51836826, and 51836851.
510(k) numbers
["K010435"]
PMA numbers
not on file
Product code
IXL
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register