FDA Medical Device Recalls (openFDA)
Ferno-Washington Inc: Poly-bags containing the strap labeled with "Adult" sticker. Each strap kit contains adjustable length leg straps, ankle wrap, a thigh pad and a heel stand. The strap kits are applied to the traction splint frame to immobilize and provide traction for a patients leg.
- Recall number
- Z-1604-2016
- Recalling firm
- Ferno-Washington Inc
- Product
- Poly-bags containing the strap labeled with "Adult" sticker. Each strap kit contains adjustable length leg straps, ankle wrap, a thigh pad and a heel stand. The strap kits are applied to the traction splint frame to immobilize and provide traction for a patients leg.
- Status
- Terminated
- Initiated
- 2016-03-10
- Posted
- not on file
- Terminated
- 2016-08-23
- Reason
- On March 9, 2016, the firm was contacted regarding the possibility of incorrect labeling on the outside poly- bag of the traction splint straps. The firm investigated and determined its supplier had labeled the outside packaging for the pediatric and adult traction splints straps incorrectly. The packaging labeled Adult contained a pediatric splint strap and the packaging labeled Pediatric conta
- Root cause
- Process control
- Action
- Ferno-Washington, Inc , has contacted all customers via telephone on March 10, 2016. All products will be replaced with new products. For further questions, please call: (937) 283-2885.
- Quantity
- 270
- Distribution
- Worldwide Distribution - US Distribution to the states of : AK ,AZ ,CA, CO, FL ,GA ,HI, IL ,IN ,KY ,MA ,MI ,MS ,NC ,NE ,NJ ,NY ,OH,ON,TX ,VA ,WI , and WV. There are no government accounts for this recall. There is one Canadian account for this recall. There are no Mexican accounts for this recall. and to the countries of : UAE, MALAYSIA, UNITED KINGDOM, SLOVAKIA, SOUTH AFRICA, QATAR and AUSTRALIA..
- Lot, serial or model codes
- Strap Kit Adult Model 444
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NZD
- Firm FEI
- 1523574
- Firm city
- Wilmington
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28