Skip to the content

FDA Medical Device Recalls (openFDA)

HeartWare Inc: HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit Product Usage: Is intended to repair the driveline cable once an electrical break has been identified. These repairs are limited to the portion of the driveline that is outside the patient boundary (exit side).

Recall number
Z-1604-2015
Recalling firm
HeartWare Inc
Product
HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit Product Usage: Is intended to repair the driveline cable once an electrical break has been identified. These repairs are limited to the portion of the driveline that is outside the patient boundary (exit side).
Status
Open, Classified
Initiated
2015-04-29
Posted
2015-05-14
Terminated
not on file
Reason
Failures of the splice repair kit if exposed to excessive force.
Root cause
Device Design
Action
HeartWare, Inc. sent an Urgent Medical Device Correction letter to all affected consignees on May 11, 2015. The letter identified the affected product, problem, and actions to be taken by the consignees. The firm requested that consignees: 1) review the notice and the Patient Communication contents, 2) forward the notice to individuals within their organization who need to be aware of the notice, 3) identify patients that are currently on support and have undergone a driveline splice repair, 4) distribute the Patient Communication to the affected patients in person, 5) continue to reinforce the messages described in the notice with patients who have experienced a splice repair during their regularly scheduled appointments, 6) complete, sign, and return the "Acknowledge and Completion Form to HeartWare within 30 days of receipt of the letter. Should any questions or concerns arise, please contact there local HeartWare representative. Their 24-hour Clinical Support personnel are also available at 1 (888) 494-6365.
Quantity
35
Distribution
Worldwide Distribution - US Nationwide in the states of FL, KY, MO, NY, TN,TX, and the countries of: Canada, Germany, Italy, Netherlands, Turkey and United Kingdom.
Lot, serial or model codes
Catalog #: ASY00116: HVAD Pump Driveline Splice Kit Lot #'s 532770, 567949, 480786, 483769, 490692, 539664, and 598189 and Catalog #: ASY00281: HVAD Pump Driveline Splicekit; Large OD Sheath Lot #'s 654090, 672401, 678800, 690423, 833963, 873159, and 880816
510(k) numbers
not on file
PMA numbers
["P100047"]
Product code
DSQ
Firm FEI
3007042319
Firm city
Miami Lakes
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register