FDA Medical Device Recalls (openFDA)
Southwest Technologies Inc: HYPOTHERMIA SLIPPERS Supportive in Cancer, packaged under the following brands: (a) Elasto-Gel(TM), SL3000 (b) Elasto-Gel(TM), SL3000LV (c) Elasto-Gel(TM), SL3001 (d) Elasto-Gel(TM), SL3007 (e) Elasto-Gel(TM), SL3008
- Recall number
- Z-1602-2018
- Recalling firm
- Southwest Technologies Inc
- Product
- HYPOTHERMIA SLIPPERS Supportive in Cancer, packaged under the following brands: (a) Elasto-Gel(TM), SL3000 (b) Elasto-Gel(TM), SL3000LV (c) Elasto-Gel(TM), SL3001 (d) Elasto-Gel(TM), SL3007 (e) Elasto-Gel(TM), SL3008
- Status
- Open, Classified
- Initiated
- 2017-12-28
- Posted
- not on file
- Terminated
- not on file
- Reason
- Hypothermia mitts, hypothermia slippers, and hypothermia caps on the market are misbranded for use during chemotherapy treatments. Additionally, Southwest Technologies, Inc. was informed of 2 events of frostbite to digits in the last year while using the hypothermia mitts.
- Root cause
- Other
- Action
- Southwest Technologies initiated their recall on December 28, 2017 by Certified Mail and expanded their recall to include international consignees on January 31, 2018 by Fed Ex. The consignees were requested to return or discard (US) and discard only (International).
- Quantity
- 12352 units
- Distribution
- worldwide
- Lot, serial or model codes
- US Distribution: (a) SL3000, Lot Numbers: 110713C 122612D 091412C 070909D 052213E 080813D 020314D 031714D 071014B 101514B 120214B 011415B 030515C 041315C 071315C 091115C 100915C 020516D 030316D 041116D 060116B 062916C 081716B 101216D OUS Distribution: (a) SL3000, Lot Numbers: 102909D 070909D 011110C 030210C 042110D 050610C 080910C 100610D 011911C 031111D 060211C 080911C 100611B 012512B 031912C 061212D 091412C 110712B 122612D 052213E 080813D 110713C 020314D 031714D 071014B 101514B 120214B 011415B 030515C 041315C 071315C 091115C 100915C 020516D 030316D 040816B 041116D 060116B 081716B 101216D 013117B 022017D 041917C (b) SL3000LV, Lot Numbers: 022117C 032817C 050417B 061917C 071217B (c) SL3001, Lot Numbers: 032410C, 110210B (d) SL3007, Lot Numbers: 080510E, 102616D (e) SL3008, Lot Numbers: 030110C
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- N/A
- Firm FEI
- 1929833
- Firm city
- Kansas City
- Firm state
- MO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28