Skip to the content

FDA Medical Device Recalls (openFDA)

Biomerieux Inc: Vitek 2 AST-YS07 REF 414 967, Fungal Susceptibility Card, 20 cards per carton, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed.

Recall number
Z-1602-2015
Recalling firm
Biomerieux Inc
Product
Vitek 2 AST-YS07 REF 414 967, Fungal Susceptibility Card, 20 cards per carton, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed.
Status
Terminated
Initiated
2015-03-27
Posted
2015-05-12
Terminated
2017-03-24
Reason
The test kits have Out-of-Range-Low (OORL) Flucytosine QC results for Candida krusei ATCC 6258
Root cause
Other
Action
The notification, consisting of the Field Safety Corrective Action (FSCA #2488) packet, was issued electronically to the subsidiaries and distributors on 3/27/15. FSCA #2509 was issued electronically to the subsidiaries and distributor on 4/13/15, which superseded FSCA #2488.
Quantity
43,535/20-card cartons
Distribution
Distributed only to foreign subsidiaries and distributors, including Canada and Mexico.
Lot, serial or model codes
lot 287303120, 02-APR-15; lot 287303122, 02-APR-15; lot 287306710, 08-MAY-15; lot 287308210, 23-MAY-15; lot 287311710, 27-JUN-15; lot 287314910, 29-JUL-15; lot 287314912, 29-JUL-15; lot 287315710, 06-AUG-15; lot 287317310, 22-AUG-15; lot 287317710, 26-AUG-15; lot 287318920, 07-SEP-15; lot 287319110, 09-SEP-15; lot 287319820, 16-SEP-15; lot 287321940, 07-OCT-15; lot 287322540, 13-OCT-15; lot 287323520, 23-OCT-15; lot 287324220, 30-OCT-15; lot 287324240, 30-OCT-15; lot 287326220, 19-NOV-15; lot 287327620, 03-DEC-15; lot 287329620, 23-DEC-15; lot 287330520, 01-JAN-16; lot 287331920, 15-JAN-16; lot 287332520, 21-JAN-16; lot 287332522, 21-JAN-16; lot 287334810, 13-FEB-16; lot 287335420, 19-FEB-16; lot 287336220, 27-FEB-16; lot 287337620, 12-MAR-16; lot 287338020, 16-MAR-16; lot 287340140, 06-APR-16; lot 287341820, 23-APR-16; lot 287341822, 23-APR-16; lot 287344620, 21-MAY-16; lot 287344622, 21-MAY-16; lot 287346710, 11-JUN-16; lot 287346720, 11-JUN-16; lot 287350120, 15-JUL-16; and lot 287350122, 15-JUL-16.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LTW
Firm FEI
1950204
Firm city
Hazelwood
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28