Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Syngo Dynamics; Kinetdx Picture Archiving and Communications System

Recall number
Z-1601-2016
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Syngo Dynamics; Kinetdx Picture Archiving and Communications System
Status
Terminated
Initiated
2016-04-14
Posted
2016-05-04
Terminated
2017-03-20
Reason
Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in carrying forward the entire section of the report including data from previous studies. This may result in the physician making analysis based on the old measurements.
Root cause
Software design
Action
Siemens mailed out a "Customer Advisory Notice" #SY009/16/P to customers informing them of the issue and what to do if it was to occur. Siemens informed customers that they are working on the "carry forward" feature enhancement and is planning to release a software update VA10B_HF06 and they should follow the instructions in the advisory notice until the update is performed.
Quantity
84 units
Distribution
Nationwide: AZ, CA, IL, IN, MA, MD, MI, MN, NJ, NY, OH, OR, PA, TN, TX, UT, WA, WI
Lot, serial or model codes
Model Numbers: 10091637, 10091673, 10091805, 10091807  Serial Numbers: 85021, 85040, 85042, 85045, 85047, 85057, 85079, 85080, 85081, 85086, 85090, 85115, 85143, 85145, 85146, 85147, 85149, 85157, 85167, 85168, 85170, 85172, 85177, 85178, 85185, 85186, 85189, 85191, 85234, 85235, 85239, 85281, 85307, 85309, 85322, 85325, 85326, 85329, 85335, 85338, 85350, 85353, 85354, 85368, 85370, 85374, 85377, 85380, 85385, 85387, 85401, 85418, 85421, 85422, 85427, 85435, 85442, 85447, 85462, 85467, 85468, 85469, 85470, 85473, 85481, 85482, 85484, 85859, 85860, 85866, 85867, 85871, 85872, 85873, 85874, 85875, 85876, 85880, 85883, 85884, 85893, 85896, 85910, 85918
510(k) numbers
["K023772"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register