FDA Medical Device Recalls (openFDA)
Nuvectra: Nuvectra, Algovita Spinal Cord Stimulation System Clinical Programmer, Model 4500. For use with Algovita Spinal Cord Stimulation Systems. The version of the software in the Clinician Programmer is v1.1.5. The Algovita Spinal Cord Stimulation system consists of a stimulator (EPG or IPG) that is physically and electrically connected to one or more leads inserted into the patient"s spinal epidural space. The lead delivers electrical stimulation originating at the stimulator with the purpose of blocking pain signals going to the patient"s brain. The stimulators are rechargeable, and the frequency of recharging is dependent on individual patient"s use of the system. The Clinician Programmer (CP) is a hand-held, touch screen device used to create and adjust stimulation parameters that are developed to optimize the therapy for each patient. The CP communicates with the stimulator via the Medical Implant Communication Service (MICS). The CP wirelessly transfers the stimulation parameters to the stimulator. In most clinical settings, the CP can be used by the Health Care Professional to program multiple patient EPGs and IPGs
- Recall number
- Z-1600-2016
- Recalling firm
- Nuvectra
- Product
- Nuvectra, Algovita Spinal Cord Stimulation System Clinical Programmer, Model 4500. For use with Algovita Spinal Cord Stimulation Systems. The version of the software in the Clinician Programmer is v1.1.5. The Algovita Spinal Cord Stimulation system consists of a stimulator (EPG or IPG) that is physically and electrically connected to one or more leads inserted into the patient"s spinal epidural space. The lead delivers electrical stimulation originating at the stimulator with the purpose of blocking pain signals going to the patient"s brain. The stimulators are rechargeable, and the frequency of recharging is dependent on individual patient"s use of the system. The Clinician Programmer (CP) is a hand-held, touch screen device used to create and adjust stimulation parameters that are developed to optimize the therapy for each patient. The CP communicates with the stimulator via the Medical Implant Communication Service (MICS). The CP wirelessly transfers the stimulation parameters to the stimulator. In most clinical settings, the CP can be used by the Health Care Professional to program multiple patient EPGs and IPGs
- Status
- Terminated
- Initiated
- 2016-04-06
- Posted
- 2016-05-03
- Terminated
- 2017-09-29
- Reason
- If used, the optional Swap feature copies programming parameters from the external pulse generator (EPG) as well as unintended calibration data to the implantable pulse generator (lPG). This may result in the user being unable to recharge the lPG. This safety notice only applies to CPs Model 4500.
- Root cause
- Software design
- Action
- Nuvectra sent an "Urgent Field Safety Notice" letter dated April 6, 2016 to customers. The letter identified the affected product, problem, recommendations and Mitigations, and actions to be taken.
- Quantity
- 17 OUS
- Distribution
- Germany
- Lot, serial or model codes
- Serial 000010 000011 FF0043 FF0045 000014 000015 000016 FF0054 FF0048 FF004B FF005D FF0064 FF0065 FF0067 FF00A9 FF00AD FF00BB Lot P178713 P185700 P187989 P195875 P213108 P257056
- 510(k) numbers
- not on file
- PMA numbers
- ["P130028"]
- Product code
- LGW
- Firm FEI
- 3010309840
- Firm city
- Blaine
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28