Baxter Healthcare Corp.: Brand Name: Infusor SV and LV Elastomeric Infusion Devices. Indicated for the intravenous administration of medications.

Recall number
Z-1600-2013
Recalling firm
Baxter Healthcare Corp.
Product
Brand Name: Infusor SV and LV Elastomeric Infusion Devices. Indicated for the intravenous administration of medications.
Status
Terminated
Initiated
2013-06-07
Posted
2013-06-27
Terminated
2014-06-04
Reason
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality. Complaint rates for ruptures have since decreased. Baxter wants to ensure that no affected product remains in the field.
Root cause
Nonconforming Material/Component
Action
An Urgent Product Recall notification was sent to all affected consignees on 06/13/2013. The letter states the affected products, problem description, hazard involved and actions to be taken by customer/user, with contact information. The letter was accompanied by a Customer Reply Form and a table of Affected Product List. The recall notification instructs customers to segregate affected product and contact Baxter for instructions for return of product.
Quantity
3,255,096 total
Distribution
Distributed Nationwide and in Puerto Rico.
Lot, serial or model codes
Product Codes: 2C1008KP, 2C1009KP, 2C1063KP, 2C1087KP, 2C1156KP, 2C1700KP, 2C1702KP  Lot Codes: 10F032 10G056 10H066 10J047 10K024 10K087 10M021 10M088 10F059 10F069 10F070 10G029 10G043 10G062 10G067 10H032 10H093 10J049 10J060 10J089 10J105 10K028 10K030 10K038 10K088 10K096 10K102 10M025 10M044 10F022 10F068 10F072 10G030 10G045 10G046 10G047 10G050 10G051 10H016 10H018 10H087 10H094 10J053 10J058 10J062 10J102 10K026 10K035 10K039 10K086 10M023 10M040 10M043 10M045 10F020 10F087 10H086 10H095 10J017 10J020 10J056 10J068 10K021 10K029 10K031 10K081 10K105 10M017 10M039 10J052 10K098 10F064 10H044 10K022 10K042 10F024 10F025 10F033 10F055 10F062 10F063 10F065 10H005 10H030 10H042 10J015 10J019 10J061 10J065 10K032 10K037 10K095 10M030 10M087 10M098 10M099 10M100  Expiration Date: 06/30/13
510(k) numbers
["K071222"]
PMA numbers
not on file
Product code
MEB
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register