Fujifilm Medical Systems U.S.A., Inc.: Smart CR Digital Computed Radiography System; Model Number: CR-IR346RU

Recall number
Z-1599-2016
Recalling firm
Fujifilm Medical Systems U.S.A., Inc.
Product
Smart CR Digital Computed Radiography System; Model Number: CR-IR346RU
Status
Terminated
Initiated
2016-04-01
Posted
not on file
Terminated
2018-07-11
Reason
Reports of failure of the device's power supply.
Root cause
Device Design
Action
Fujifilm sent an Medical Device Correction letter dated March 22, 2016, via FedEx to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Instructions for turning the unit off when not in use as an interim action. FMSU will conduct a field replacement of the power supply in all installed units. For further questions, please call (888) 385-4633.
Quantity
2,062 devices
Distribution
Nationwide Distribution.
Lot, serial or model codes
All serial numbers.
510(k) numbers
["K944046"]
PMA numbers
not on file
Product code
IXW
Firm FEI
1000513161
Firm city
Stamford
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register