Skip to the content

FDA Medical Device Recalls (openFDA)

ConBio, a CynoSure Company: Multilite Dye Handpieces, an optional accessory of the RevLite SI Laser. The Revlite SI laser is intended for General Dermatology, Dermatologic and General Surgical Procedures. Which includes treatments such as tattoo removal, pigmented lesion removal, removal or lightening of hair, skin resurfacing and reduction of red pigmentation in hypertropic and keloid scars The two Multilite Dye Handpieces (HPs) are optional handpieces that can be used with the Revlite SI Laser. Manufactured by: ConBio, A Cynosure Company 47733 Fremont blvd Fremont, CA 94538

Recall number
Z-1599-2013
Recalling firm
ConBio, a CynoSure Company
Product
Multilite Dye Handpieces, an optional accessory of the RevLite SI Laser. The Revlite SI laser is intended for General Dermatology, Dermatologic and General Surgical Procedures. Which includes treatments such as tattoo removal, pigmented lesion removal, removal or lightening of hair, skin resurfacing and reduction of red pigmentation in hypertropic and keloid scars The two Multilite Dye Handpieces (HPs) are optional handpieces that can be used with the Revlite SI Laser. Manufactured by: ConBio, A Cynosure Company 47733 Fremont blvd Fremont, CA 94538
Status
Terminated
Initiated
2013-05-09
Posted
2013-06-26
Terminated
2013-06-26
Reason
An inconsistency was identified related to electronic display identification of the optional Multilite Dye Handpieces of the RevLite SI Laser system that may confuse user.
Root cause
Nonconforming Material/Component
Action
The firm, Cynosure Inc., sent a "MEDICAL DEVICE FIELD ACTIONS NOTIFICATION" letter dated May 9, 2013 to its customers via UPS next day. The letter described the product, problem and actions to be taken. The customers were instructed to immediately stop using the Revlite SI Multilite Dye Handpieces upon receipt of this letter and complete and return the enclosed Field Corrective Action Acknowledgement Form via e-mail to: mflynn@cynosure.com or FAX to: 978-256-6556. A Cynosure representative will be contacting customers to schedule and arrange an exchange of the affected handpieces. If additional information is required please contact VP of Regulatory Affairs and Quality Systems at 978-367-2350 or via email at ikulinets@cynosure.com.
Quantity
12 (6 of model 659-1050; 6 of model 659-1100)
Distribution
Worldwide distribution: USA (nationwide) including states of: IL, ME, NM and VA; and countries of: Australia and Chile.
Lot, serial or model codes
659-1050-585 nm Handpiece; Manufacturer numbers: 659-1050 / C13B1141 659-1050 / D13C1151 659-1050 / D13C1157 659-1050 / D13C1158 659-1050 / D13C1160 659-1050 / D13C1154 Model 659-1100 - 650 nm Handpiece; manufacturer numbers: 659-1100 / C13B1141 659-1100 / D13C1151 659-1100 / D13C1157 659-1100 / D13C1158 659-1100 / D13C1160 659-1100 / D13D1165.
510(k) numbers
["K103118"]
PMA numbers
not on file
Product code
GEX
Firm FEI
1000123592
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register