Skip to the content

FDA Medical Device Recalls (openFDA)

Spinal Elements, Inc: Spinal Elements 5.5 x 45MM Mercury lordosed rod. A component of the Mercury Spinal System, comprised of a variety of screws, hooks, rods, connectors, and staples that are used for attachment to the non-cervical spine. Rods span the distance between screws and hooks and achieve fixation by the mechanical joining of the rods with the screws or hooks. Lordosed rods are pre-curved to better fit the patient anatomy.

Recall number
Z-1597-2015
Recalling firm
Spinal Elements, Inc
Product
Spinal Elements 5.5 x 45MM Mercury lordosed rod. A component of the Mercury Spinal System, comprised of a variety of screws, hooks, rods, connectors, and staples that are used for attachment to the non-cervical spine. Rods span the distance between screws and hooks and achieve fixation by the mechanical joining of the rods with the screws or hooks. Lordosed rods are pre-curved to better fit the patient anatomy.
Status
Terminated
Initiated
2015-04-09
Posted
2015-05-11
Terminated
2015-06-25
Reason
Some rods may not meet the appropriate traceability requirements including the lot number. Some rods were found to not include the product identification number (product number and lot number) marked directly on the rod. The lack of marking interrupts the ability to maintain traceability of the device.
Root cause
Under Investigation by firm
Action
The firm sent customer notification letters on 04/09/15. In this letter the firm requested that all affected lots be returned, and replacement inventory will be made available. The firm requested that customers take the following actions: 1. Immediately examine inventory for part number 60100-045. 2. Immediately discontinue the use and/or distribution of the affected lots 3. Send all affected product back to: Spinal Elements, Inc. ATTN: Julie Lamothe 3115 Melrose Drive Carlsbad, CA 92010 4. If further distributed the product to please identify the customers and notify them as once. 5. Complete and return the enclosed "Customer Response Form." Any questions the firm states to contact the Customer Service department at 1.877.SPINAL5 Ext. 213.
Quantity
588
Distribution
Nationwide Distribution including the states of GA, CA, TX, CO, IN, NC, FL, UT, TN, MO, MI, PA, OH, and OK.
Lot, serial or model codes
Part Number: 60100-045 Lot Numbers: 140086, 140449, 141013
510(k) numbers
["K071914"]
PMA numbers
not on file
Product code
NKB
Firm FEI
3004893332
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register