RF Technologies, Inc.: Code Alert Advanced 4-Way Care Solution, CA520 System, PN 9600-1600, RF Technologies. This product is used to alert caregivers when a patient at risk for falls is leaving a bed or chair.

Recall number
Z-1597-2013
Recalling firm
RF Technologies, Inc.
Product
Code Alert Advanced 4-Way Care Solution, CA520 System, PN 9600-1600, RF Technologies. This product is used to alert caregivers when a patient at risk for falls is leaving a bed or chair.
Status
Terminated
Initiated
2013-05-17
Posted
2013-06-26
Terminated
2013-12-03
Reason
RF Technologies, Inc. has initiated a recall involving CA520 Code Alert Integrated Care Management Advanced 4-Way Care Solution . Due to a firmware error, the alarm may not be transmitted to the system server immediately.
Root cause
Nonconforming Material/Component
Action
Consignees were sent on 5/17/2013 a "Medical Device Recall" letter dated 3 May 2013. The letter described the problem and the product involved in the recall. The letter advised consignees to discontinue use of the device and return to RF Technologies.
Quantity
167
Distribution
Nationwide Distribution in US including the states of AK, AZ, CA, GA, IA, KY, MT, NE, NJ, ND, and WI.
Lot, serial or model codes
all codes
510(k) numbers
not on file
PMA numbers
not on file
Product code
KMI
Firm FEI
3002286578
Firm city
Brookfield
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register