Skip to the content

FDA Medical Device Recalls (openFDA)

Hospira Inc: Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser

Recall number
Z-1596-2011
Recalling firm
Hospira Inc
Product
Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser
Status
Terminated
Initiated
2010-10-27
Posted
2011-03-08
Terminated
2013-11-07
Reason
Complaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords.
Root cause
Component design/selection
Action
Hospira Recall notification letters were sent by Federal Express on November 1, 2010.
Quantity
32,376 power cords
Distribution
Nationwide distribution and to Australia, Canada, and Malaysia.
Lot, serial or model codes
Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser List number configurations: 16026-04-01/03/04, 51/52,. 53/54, 79/80, 81/82, 83/84/ 87/88; 16027-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88; 16026-13-25/26, 53/54; 16027-13-25/26, 53/54.
510(k) numbers
["K051550"]
PMA numbers
not on file
Product code
FRN
Firm FEI
2921482
Firm city
Morgan Hill
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register