FDA Medical Device Recalls (openFDA)
Cook Inc.: StoneBreaker Probe Cap, Catalog number:SBA-PC; GPN: G52601; UDI:(01)00827002526013
- Recall number
- Z-1593-2018
- Recalling firm
- Cook Inc.
- Product
- StoneBreaker Probe Cap, Catalog number:SBA-PC; GPN: G52601; UDI:(01)00827002526013
- Status
- Terminated
- Initiated
- 2018-01-31
- Posted
- not on file
- Terminated
- 2020-04-17
- Reason
- The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
- Root cause
- Error in labeling
- Action
- On March 8, 2018 Cook Medical sent customers notification of an Urgent Medical Device Recall . Customers were instructed to complete the following actions: 1. Examine inventory immediately to determine if you have affected product, and quarantine affected product(s). Immediately cease all distribution and use of these products. 2.Return affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form. Used devices being returned should be double-bagged, placed inside an outer puncture free package with sufficient cushioning material to prevent movement between the secondary container and the outer package. An itemized list of the components must be placed in the container and the outer package marked Used Medical Device. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product on hand, you must still complete the Acknowledgement and Receipt Form and return it via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. Immediately report adverse events to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Customers with medical questions or concerns can contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235
- Quantity
- 370
- Distribution
- US Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, UA, UY, XK, ZA
- Lot, serial or model codes
- All Lots
- 510(k) numbers
- ["K062475"]
- PMA numbers
- not on file
- Product code
- FFK
- Firm FEI
- 1820334
- Firm city
- Bloomington
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28