Carestream Health Inc: CARESTREAM Image Suite V3: MINI-PACS/F IMG ST/CLASSIC: REF/Catalog # 1036490; MINI-PACS/F IMG ST/POC&VITA: REF/Catalog # 1036508; MINI-PACS/F IMG ST/CLASSIC/INDIA: REF/Catalog # 1036417; MINI-PACS/F IMG ST/ POC&VITA /INDIA: REF/Catalog # 1036425 -- Made in USA by: Carestream Health, Inc. 150 Verona Street, Rochester, NY 14608 --- CLASSIFICATION NAME: System, Image Processing, Radiological The Carestream Image Suite System is an image management system whose intended use is to receive, process, review, display, print and archive images and data from CR and DR modalities. This excludes mammography applications in the United States.
- Recall number
- Z-1593-2016
- Recalling firm
- Carestream Health Inc
- Product
- CARESTREAM Image Suite V3: MINI-PACS/F IMG ST/CLASSIC: REF/Catalog # 1036490; MINI-PACS/F IMG ST/POC&VITA: REF/Catalog # 1036508; MINI-PACS/F IMG ST/CLASSIC/INDIA: REF/Catalog # 1036417; MINI-PACS/F IMG ST/ POC&VITA /INDIA: REF/Catalog # 1036425 -- Made in USA by: Carestream Health, Inc. 150 Verona Street, Rochester, NY 14608 --- CLASSIFICATION NAME: System, Image Processing, Radiological The Carestream Image Suite System is an image management system whose intended use is to receive, process, review, display, print and archive images and data from CR and DR modalities. This excludes mammography applications in the United States.
- Status
- Terminated
- Initiated
- 2016-04-15
- Posted
- 2016-05-02
- Terminated
- 2017-03-16
- Reason
- Carestream Health received a complaint related to CARESTREAM Image Suite 4 from a foreign hospital stating that the annotation on the overlay is displayed as "L (Left)", when it should be "R(Right)".
- Root cause
- Software design
- Action
- Carestream Health sent an URGENT MEDICAL DEVICE CORRECTION letter dated April 6, 2016, and Consignee Notification Acknowledgement form to the US consignees on April 15, 2016, via Certified Mail, Return Receipt Requested. Foreign consignees are being notified based on the individual country regulations. Customers who have this Image Suite configuration but do not use it for CT or MR Imaging do not need to take any action. Customers who use the product for CT or MR Images are asked to make users aware of the problem and take additional steps to check correct orientation of any markers, until the software is updated. Customers with questions or concerns were instructed to contact the Carestream Customer Care Center in the US at 1-800-328-2910. Customers outside the US were instructed to contact their local Service Support number. For questions regarding this recall call 1-800-328-2917.
- Quantity
- US: 50 units; Foreign: 2499 units
- Distribution
- NY and OH for further distribution nationwide; Foreign distribution to the following countries: Afghanistan, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia & Herzgovena, Brazil, Bulgaria, Cambodia, Canada, Chad, Chile, China, Colombia, Croatia, Czech Rep., Djibouti, France, Germany, Greece, Hungary, India, Indonesia, Iran, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon, Liechtenstein, Lithuania, Macao, Malaysia, Mexico, Montenegro, Myanmar (Burma), Nepal, Netherlands, New Zealand, Niger, Norway, Oman, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Scotland, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Taiwan, thailand, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam.
- Lot, serial or model codes
- Equipment ID Numbers: US units: 8501-2596, 85018860, 8501-0954, 5241-2596, 5241-2546, 5241-2713, 5296-3447, 52414419, 52414455, 5241-2640, 5241-2716, 54838552, 5483-8502, 8501-4665, 52408158, 85014687, 54839128, 5483-9212, 52412539, 52414661, 52962897, 4745-3986, 5241-2337, 85012262, 8501-5898, 8501-2602, 54838630, 52412670, 5296-3450, 5296-3449, 54838554, 52408157, 54838618, 5241-4547, 54842652, 52420675, 8501-2861, 52414685, 52414751, 5484-3086, 54838598, 52412367, 52414417, 54839211, 54838616, 4745-5991, 52412243, 54842667, 52417149, 52412340; -- Equipment ID Numbers for Foreign Units: Please contact CDRH Recall Group for list of affected Equipment ID Numbers.
- 510(k) numbers
- ["K100094"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 1315356
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28