FDA Medical Device Recalls (openFDA)
Abbott Spine: PathFinder Spinal Fixation System T-Handle Driver; Part 2151-7 (13" shaft); Non-sterile, Reusable Instrument; manufactured by Abbott Spine, Austin, TX 78727.
- Recall number
- Z-1593-2008
- Recalling firm
- Abbott Spine
- Product
- PathFinder Spinal Fixation System T-Handle Driver; Part 2151-7 (13" shaft); Non-sterile, Reusable Instrument; manufactured by Abbott Spine, Austin, TX 78727.
- Status
- Terminated
- Initiated
- 2008-03-24
- Posted
- 2008-09-01
- Terminated
- 2009-12-02
- Reason
- Reports received regarding disassembly of T-Handle Driver components during surgery dropping the closure top into the surgical site.
- Root cause
- Error in labeling
- Action
- Firm sent notification letters to sales agents via letter on 03/24/08. Enclosed with the notification letter, the firm provided updated product inserts for the InCompass and PathFinder Spinal Fixation System (Cat. #2199-0002-LB Rev. H and #1199-0002-LB Rev. H, respectively), as well as the "Disassembly and Assembly Instructions for PathFinder & InCompass T-Handle Drivers" (Cat #1199-0040-MKC Rev. A). Sales Agents instructed to distribute updated inserts and instructions to all locations that handle the drivers for cleaning and to destroy all copies of old inserts. If you have questions or have not received the recall notice from the firm, contact Kirstin Widding at 512-533-1852.
- Quantity
- 224 units.
- Distribution
- Nationwide Distribution --- including states of AR, AZ, CA, CT, CO, FL, IL, LA, MD, MI, MO, MN, MT, NC, NM, NJ, NY, OH, OK, OR, PA, SD, TX, VA and WI.
- Lot, serial or model codes
- All lots.
- 510(k) numbers
- ["K030625","K071174"]
- PMA numbers
- not on file
- Product code
- MNI
- Firm FEI
- 3000718496
- Firm city
- Austin
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as