Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Spine: PathFinder Spinal Fixation System T-Handle Driver; Part 2151-7 (13" shaft); Non-sterile, Reusable Instrument; manufactured by Abbott Spine, Austin, TX 78727.

Recall number
Z-1593-2008
Recalling firm
Abbott Spine
Product
PathFinder Spinal Fixation System T-Handle Driver; Part 2151-7 (13" shaft); Non-sterile, Reusable Instrument; manufactured by Abbott Spine, Austin, TX 78727.
Status
Terminated
Initiated
2008-03-24
Posted
2008-09-01
Terminated
2009-12-02
Reason
Reports received regarding disassembly of T-Handle Driver components during surgery dropping the closure top into the surgical site.
Root cause
Error in labeling
Action
Firm sent notification letters to sales agents via letter on 03/24/08. Enclosed with the notification letter, the firm provided updated product inserts for the InCompass and PathFinder Spinal Fixation System (Cat. #2199-0002-LB Rev. H and #1199-0002-LB Rev. H, respectively), as well as the "Disassembly and Assembly Instructions for PathFinder & InCompass T-Handle Drivers" (Cat #1199-0040-MKC Rev. A). Sales Agents instructed to distribute updated inserts and instructions to all locations that handle the drivers for cleaning and to destroy all copies of old inserts. If you have questions or have not received the recall notice from the firm, contact Kirstin Widding at 512-533-1852.
Quantity
224 units.
Distribution
Nationwide Distribution --- including states of AR, AZ, CA, CT, CO, FL, IL, LA, MD, MI, MO, MN, MT, NC, NM, NJ, NY, OH, OK, OR, PA, SD, TX, VA and WI.
Lot, serial or model codes
All lots.
510(k) numbers
["K030625","K071174"]
PMA numbers
not on file
Product code
MNI
Firm FEI
3000718496
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register