FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corp.: Epidural Infusion Pumps sold under the following labels; a) Baxter 6060E Epidural Infusion Pump, product code 2L9351, 2L9351P and 2L9351R; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore b) Sabratek 6060 Epidural Pain Management Pump, product codes 606000-40E and 606000-40EL; Manufactured by Sabratek Corp., Skokie, IL 60076, which was purchased by Baxter Healthcare Corp., Deerfield, IL in January 2000.
- Recall number
- Z-1592-05
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Epidural Infusion Pumps sold under the following labels; a) Baxter 6060E Epidural Infusion Pump, product code 2L9351, 2L9351P and 2L9351R; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore b) Sabratek 6060 Epidural Pain Management Pump, product codes 606000-40E and 606000-40EL; Manufactured by Sabratek Corp., Skokie, IL 60076, which was purchased by Baxter Healthcare Corp., Deerfield, IL in January 2000.
- Status
- Terminated
- Initiated
- 2005-08-17
- Posted
- 2005-09-28
- Terminated
- 2007-06-19
- Reason
- Potential overinfusion in the Auto-Ramp mode with software versions 3.00 or higher
- Root cause
- Other
- Action
- Baxter sent the 8/17/05 Urgent Product Recall letter to all Baxter 6060E Infusion Pump customers to inform them of a software anomaly present in software versions 3.00 and higher, which allows the pump to overinfuse in the Auto-Ramp mode. The end users were also informed that these epidural pumps, identified by the yellow housing, can no longer be supported due to a lack of availability of spare parts and Baxter is unable to replace the software with an alternate version. As a result, Baxter requested the accounts to remove the pumps from service and return them to Baxter. The accounts were requested to call Baxter''s Medication Delivery Services at 1-800-843-7867 to arrange for the return of the units and replacement with 6060 Multi-Therapy Infusion Pumps.
- Quantity
- not on file
- Distribution
- Nationwide including Puerto Rico, and intenationally to Canada, China, Belgium, Hong Kong and Saudi Arabia
- Lot, serial or model codes
- product codes 2L9351, 2L9351P, 2L9351R, 606000-40E and 606000-40EL: all serial numbers
- 510(k) numbers
- ["K002679","K941984"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 1416980
- Firm city
- Round Lake
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28