Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Epidural Infusion Pumps sold under the following labels; a) Baxter 6060E Epidural Infusion Pump, product code 2L9351, 2L9351P and 2L9351R; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore b) Sabratek 6060 Epidural Pain Management Pump, product codes 606000-40E and 606000-40EL; Manufactured by Sabratek Corp., Skokie, IL 60076, which was purchased by Baxter Healthcare Corp., Deerfield, IL in January 2000.

Recall number
Z-1592-05
Recalling firm
Baxter Healthcare Corp.
Product
Epidural Infusion Pumps sold under the following labels; a) Baxter 6060E Epidural Infusion Pump, product code 2L9351, 2L9351P and 2L9351R; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore b) Sabratek 6060 Epidural Pain Management Pump, product codes 606000-40E and 606000-40EL; Manufactured by Sabratek Corp., Skokie, IL 60076, which was purchased by Baxter Healthcare Corp., Deerfield, IL in January 2000.
Status
Terminated
Initiated
2005-08-17
Posted
2005-09-28
Terminated
2007-06-19
Reason
Potential overinfusion in the Auto-Ramp mode with software versions 3.00 or higher
Root cause
Other
Action
Baxter sent the 8/17/05 Urgent Product Recall letter to all Baxter 6060E Infusion Pump customers to inform them of a software anomaly present in software versions 3.00 and higher, which allows the pump to overinfuse in the Auto-Ramp mode. The end users were also informed that these epidural pumps, identified by the yellow housing, can no longer be supported due to a lack of availability of spare parts and Baxter is unable to replace the software with an alternate version. As a result, Baxter requested the accounts to remove the pumps from service and return them to Baxter. The accounts were requested to call Baxter''s Medication Delivery Services at 1-800-843-7867 to arrange for the return of the units and replacement with 6060 Multi-Therapy Infusion Pumps.
Quantity
not on file
Distribution
Nationwide including Puerto Rico, and intenationally to Canada, China, Belgium, Hong Kong and Saudi Arabia
Lot, serial or model codes
product codes 2L9351, 2L9351P, 2L9351R, 606000-40E and 606000-40EL: all serial numbers
510(k) numbers
["K002679","K941984"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register