Skip to the content

FDA Medical Device Recalls (openFDA)

Abiomed, Inc.: Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.

Recall number
Z-1591-2026
Recalling firm
Abiomed, Inc.
Product
Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.
Status
Open, Classified
Initiated
2026-02-16
Posted
2026-03-27
Terminated
not on file
Reason
Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
Root cause
Software design
Action
Abiomed notified international consignees on about 02/16/2026 via emailed CRITICAL PRODUCT CORRECTION letter. Consignees were instructed to cease use of the Automated Impella Controller (AIC) with an Impella 5.5 S1 (product code 0550-0007) unless AIC software is updated to v10.1 or above, identify current software in the lower left corner of the initial AIC display screen, contact your local Abiomed representative to schedule your software update, review and complete the customer response form provided and post this notification in a visible area for awareness. Consignees were also instructed to forward this notification to anyone within the facility that should be aware and forward the notification to consignees if further distributed.
Quantity
62 units
Distribution
International distribution in the country of Australia, Canada, France, Germany, Kuwait.
Lot, serial or model codes
All products below with software version 8.2.2 through 8.5 are affected. 1. Product Code: 0042-0000-CA; GTIN: 00813502011272; Serial Numbers: IC2035, IC9416. 2. Product Code: 0042-0000-EU; GTIN: 00813502011289; Serial Numbers: IC1677, IC3049, IC3099, IC3102, IC4358, IC4329, IC5774, IC5757, IC6580, IC6581. 3. Product Code: 0042-0000-UK; GTIN: 00813502011296; Serial Numbers: IC2023, IC1910, IC2704, IC3325, IC3312. 4. Product Code: 0042-0000-US; GTIN: 00813502010022; Serial Numbers: IC1140. 5. Product Code: 0042-0010-EU; GTIN: 00813502010961; Serial Numbers: IC5403, IC5404, IC5635, IC5636, IC5666, IC5667, IC5801, IC5904, IC5918, IC5921, IC6421, IC6422, IC8683. 6. Product Code: 0042-0010-UK; GTIN: 00813502010978; Serial Numbers: IC8101, IC8102, IC8103, IC8104, IC8544, IC9446, IC9449, IC11440. 7. Product Code: 0042-0040-CA; GTIN: 00813502012958; Serial Numbers: IC10306, IC10312, IC12604, IC12891, IC12895, IC12892, IC12605, IC12894. 8. Product Code: 0042-0040-EU; GTIN: 00813502012217; Serial Numbers: IC9014, IC9022, IC9429, IC10408, IC10425. 9. Product Code: 0042-0040-AU; GTIN: 00813502012972; Serial Numbers: IC10163, IC10387, IC11546, IC11566, IC11490, IC11544, IC12008, IC12054, IC11472, IC12151.
510(k) numbers
not on file
PMA numbers
["P140003"]
Product code
OZD
Firm FEI
1220648
Firm city
Danvers
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register