FDA Medical Device Recalls (openFDA)
Medtronic Sofamor Danek USA Inc: INFUSE(R) Bone Graft, LARGE KIT, REF 7510600, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic
- Recall number
- Z-1591-2013
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- INFUSE(R) Bone Graft, LARGE KIT, REF 7510600, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic
- Status
- Terminated
- Initiated
- 2013-05-28
- Posted
- 2013-06-25
- Terminated
- 2014-05-13
- Reason
- Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.
- Root cause
- Nonconforming Material/Component
- Action
- Medtronic on behalf of Integra LifeSciences Corporation sent an Urgent Voluntary Removal (Recall) letter on May 24, 2013, directly to all Risk Managers and Sales Representatives, with Recall Questionnaires and Confirmation forms included. Foreign consignees were notified by email on May 28, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remove the product and discontinue use. The letter should be forwarded to surgeons or attending physicians who have used the affected products. Customers were instructed to use the attached questionnaire to account for product in their possession, document which physicians they notified and report any patient injuries that may have occurred related to the use of the affected product by calling the phone number in the questionnaire. Customers were asked to return the questionnaire using the enclosed self-addressed stamped envelope, via their Medtronic Sales Representative or via fax to 763-367-8134 For questions regarding this recall 901-399-3182. Customers were instructed to contact their Medtronic Sales Representative to coordinate the return and replenishment of affected product . Customers with questions should call 800-876-3133, ext 3197.
- Quantity
- 4885 units
- Distribution
- Worldwide distribution: USA (nationwide including Puerto Rico) and Internationally to Australia, Brazil, Canada, Chile, Hong Kong, India, Mexico, Netherlands, , New South Wales, Panama, and Paraguay.
- Lot, serial or model codes
- Lots: M111052AAE, M111052AAN, M111052AAS, M111052AAD, M111059AAA
- 510(k) numbers
- not on file
- PMA numbers
- ["P000058"]
- Product code
- NEK
- Firm FEI
- 1000477302
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28