Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical MD, Inc.: Deltec Cozmo Insulin Pump, model 21-1800. This is similar to the currently distributed Deltec Cozmo Insulin Pump Model 1700, with the exception of updated software that enhanced certain features and added a food data base to allow users to better manage dietary requirements in conjunction with the use of their insulin pump. Reorder Numbers: 21-1801-81 21-1803-81 21-1804-81 21-1805-49/51/81 21-1806-49/51/81 21-1807-49/51/81 21-1815-01/50 21-1816-01/50 21-1817-01/50

Recall number
Z-1591-2008
Recalling firm
Smiths Medical MD, Inc.
Product
Deltec Cozmo Insulin Pump, model 21-1800. This is similar to the currently distributed Deltec Cozmo Insulin Pump Model 1700, with the exception of updated software that enhanced certain features and added a food data base to allow users to better manage dietary requirements in conjunction with the use of their insulin pump. Reorder Numbers: 21-1801-81 21-1803-81 21-1804-81 21-1805-49/51/81 21-1806-49/51/81 21-1807-49/51/81 21-1815-01/50 21-1816-01/50 21-1817-01/50
Status
Terminated
Initiated
2008-02-22
Posted
2008-09-01
Terminated
2011-01-19
Reason
Smith Medical has received reports of Deltec Cozmo Insulin Pumps powering down without an alarm. Investigation concluded that if the battery cap is not fully tightened, the Pump may power down and a brief "chirp" will sound. This may occur if the user has not sufficiently tightened the battery cap or if the battery cap is damaged. If a patient is asleep or does not regularly check their blood gl
Root cause
Device Design
Action
Consignees and the USA demonstration pump physicians were mailed an "Urgent Medical Device Correction" letter on 2/22/08. The letter stated Details on Affected devices, Description of the problem, Advice on action to be taken by the user, Additional Information and New Warning. These consignees will be required to return the Correction Confirmation Form. The 17 countries with international mailing addresses were sent the Urgent Medical Device Correction letter (Urgent Field Safety Notice) via the Smiths Medical International Regulatory department based in Watford UK. The Smiths Medical Distributors and Affiliates will be required to mail the Safety Notice to their direct users. The USA distributors were mailed the Urgent Medical Device correction letter with a cover letter. These consignees will be required to add the letter to in-house inventory. The USA prescribing physician and the the USA Managed Care Accounts (Insurance Companies) were mailed the Urgent Medical Device Correction letter with a cover letter. These consignees will not be required to respond and no follow-up will be done by Smiths Medical. Contact Mike Herbert at 651-628-7049 if you have any questions or did not receive the recall information from the firm.
Quantity
not on file
Distribution
Worldwide Distribution --- USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of AU, CA, CL, CN, DE, EG, FR, GB, GR, IL, KW, MX, NL, NZ, ON, PL, SA, SE, TR, UK.
Lot, serial or model codes
all serial numbers
510(k) numbers
["K020655","K031738","K062323"]
PMA numbers
not on file
Product code
LZG
Firm FEI
2183502
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register