Philips North America: Philips Incisive CT -Computed Tomography X-Ray System Model Number: 72813

Recall number
Z-1586-2023
Recalling firm
Philips North America
Product
Philips Incisive CT -Computed Tomography X-Ray System Model Number: 72813
Status
Open, Classified
Initiated
2023-03-20
Posted
2023-05-12
Terminated
not on file
Reason
Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury
Root cause
Under Investigation by firm
Action
Philips Medical issued Urgent Medical Device Correction Letter Multi-Function Switch Unload Pedal Foot Entrapment on 3/20/23 to US consignees via Certified United States Postal Service (USPS) with delivery confirmation. International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take: Immediately stop using the Unload Pedal of the Foot Switch until Philips installs a solution on your system. " Instead of using the Unload Pedal, please use the Unload Patient function on the Gantry Control Panels, or use other Table In/Out, Up/Down functions on the Gantry Control Panels, CT Scan Control Box, or Hardware lnterventional Controls to fulfill the normal patient unloading function, as instructed in the IFU. " You may continue to use Load Pedal, and Free-Float Pedal of Foot Switch, as they are not affected by this Urgent Medical Device Correction Letter. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the solution to resolve this issue (refer to FCO72800794). If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377).
Quantity
3 units US ;2 units OUS
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, MI, MO, NY, TN, TX, VT and the countries of Australia, Chile, Costa Rica, Czech Republic, Dominican, Republic Germany, India, Israel, Japan, Lithuania, Philippines, Poland, Sint Maarten (Dutch part), South Africa, Thailand, United Kingdom, Utd. Arab. Emir.
Lot, serial or model codes
UDI-DI:(01)00884838085015(21) + Serial number US Serial Numbers: 33041, 500301, 500252. OUS Serial Numbers: 33040, 500408
510(k) numbers
["K180015","K212441"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3006648320
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register