FDA Medical Device Recalls (openFDA)
Conavi Medical Inc.: Brand Name: Novasight Hybrid System Product Name: Novasight Hybrid Catheter Model/Catalog Number: TA-06-0001 Product Description: Common Name: Diagnostic intravascular catheter Description: The Novasight Hybrid catheter is a 3Fr single use, sterile device, which will be able to perform transluminal ultrasound and optical coherence tomography imaging of adults. The catheter is capable of real-time 2D side viewing ultrasound and optical imaging acquired simultaneously providing precisely co-registered images with the ability to produce rapidly post processed rendered longitudinal images of the vessel. Packaging: The Novasight Hybrid Catheter is packaged in a packaging system including the protective packaging and the sterile barrier system.
- Recall number
- Z-1585-2025
- Recalling firm
- Conavi Medical Inc.
- Product
- Brand Name: Novasight Hybrid System Product Name: Novasight Hybrid Catheter Model/Catalog Number: TA-06-0001 Product Description: Common Name: Diagnostic intravascular catheter Description: The Novasight Hybrid catheter is a 3Fr single use, sterile device, which will be able to perform transluminal ultrasound and optical coherence tomography imaging of adults. The catheter is capable of real-time 2D side viewing ultrasound and optical imaging acquired simultaneously providing precisely co-registered images with the ability to produce rapidly post processed rendered longitudinal images of the vessel. Packaging: The Novasight Hybrid Catheter is packaged in a packaging system including the protective packaging and the sterile barrier system.
- Status
- Open, Classified
- Initiated
- 2025-03-12
- Posted
- 2025-04-23
- Terminated
- not on file
- Reason
- Due to manufacturing issues there is a potential for the catheter sheath to detach.
- Root cause
- Process control
- Action
- On 03/12/2025, the firm sent via email an "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" Letter informing customer that the firm has become aware that the sheath of the Novasight Hybrid catheter detached and left in the patient's body during a clinical case but was retrieved during the procedure. Customers are instructed to: Customers shall stop/cease use of the affected the Novasight Hybrid catheters and remove them from service and return immediately to Conavi Medical Inc. This applies to all Novasight Hybrid catheters shipped to U.S.A. regardless of expiry dates. Acknowledge receipt of this recall notification by completing and returning the "Acknowledgement" section of this notification by email to: julie@conavi.com. For questions or assistance - Contact information: Director of Product Management and Marketing Telephone # (416)-483-0100 Ext 109, Email: julie@conavi.com Days/Hours Available for calls: Monday through Friday, 9:00 AM to 4:30 PM Eastern Time.
- Quantity
- 685
- Distribution
- US distribution to states of: GA, NY, NJ, OH; and OUS (Foreign) to countries of: Canada and China
- Lot, serial or model codes
- Model No: TA-06-0001 UDI-DI: (01)00628055603054 Lot number: 230902; Expiration date: 2024-06; Lot number: 240202; Expiration date: 2025-03; Lot number: 240302; Expiration date: 2025-04; Lot number: 240402; Expiration date: 2025-05; Lot number: 240502; Expiration date: 2025-06; and all catheter lots which were expired (See attachment 2a, 2b and 2c for details).
- 510(k) numbers
- ["K172258"]
- PMA numbers
- not on file
- Product code
- OBJ
- Firm FEI
- 3012198618
- Firm city
- Toronto
- Firm state
- not on file
- Firm country
- Canada
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28