Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems Nederland B.V.: Zenition 70. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.

Recall number
Z-1583-2025
Recalling firm
Philips Medical Systems Nederland B.V.
Product
Zenition 70. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.
Status
Open, Classified
Initiated
2023-07-20
Posted
2025-04-21
Terminated
not on file
Reason
Philips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch.
Root cause
Software design
Action
Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide the following actions that should be taken to prevent risks to patients: Keep the Electronic Product Radiation Correction letter with the documentation of the system. Where possible, place the Wireless Foot Switch within a 3 meter range from the C-arm stand to minimize as much as possible the potential delay of 72 milliseconds. Circulate this Electronic Product Radiation Correction letter to all users of the system so that they are aware of the issue. Philips will contact customers to schedule a visit to arrange for a Field Service Engineer visit to update the Software free of charge. (Philips reference 2024-IGT-PUN-007).
Quantity
12 systems
Distribution
US Nationwide distribution
Lot, serial or model codes
Zenition 70
510(k) numbers
["K183040","K212813"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3000976525
Firm city
Eindhoven
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register